LXATC-01 Infrared Body Temperature
Screening Camera
Suggested Procedures for High Temperature Detection:
1. Person with high temperature should step aside and proceed to a
holding or isolation area
2. Temperature may need to be verified using an alternative medical
thermometer.
3. If high temperature is persistent, appropriate measures should be
taken immediately
Common Trouble Shooting:
Troubles
Possible Cause
Possible Reason
Treatment
Inaccurate
Temperature
measurement
LED shows
uncertain
temperature when
testing
Ambient
temperature is too
low or too high
Test in indoor /
windless environment
and ensure the
ambient is 16°C to 28°C
Device always
alert or
unable to
measure
temperature
The two warning
lights above the
device keep flashing
The device is facing
an overheated
objects or the
sensor is exposed
directly to sunlight
1.
Remove the
overheated object
in front of the
device
2.
Avoid direct
sunlight to the
sensor
Device does not
respond during the
body temperature
measurement
The device is too far
from subject
Adjust distance
between subject and
camera to
approximately 1 meter
Device cannot
connect to the
network
Green light at the
end of the network
cable is off
Network cable
connection failure
Check router, network
switch and network
cable
Computer
programming tool
cannot detect the
device
Network problem,
the device did not
get the IP address
correctly
Need to enable router
DHCP function