LXATC-01 Infrared Body Temperature
Screening Camera
Temperature Calibration Setting:
1.
Select Work Mode and click “Set Mode” button to set.
1.1 Calibration Mode
– Click when using black body / human body
to calibrate the device (this is a factory procedure)
2. Set the calibration temperature.
2.1 Black Body Calibration - Set the black body temperature
to 36.5
°C. After the black body temperature stabilizes, this
temperature can be used as the calibration temperature.
2.2 Human Body Calibration - Use a reliable forehead
thermometer to get the average of human body temperature
at least 3X, and use the average as the calibration temperature.
3. Set the human body / Black body with in 1 meter in front of the
thermometer camera and click Calib button to start the calibration.
Note: Human and black body should not move during the calibration.
4.
At the start of calibration, the device will prompt a “DEE” sound and
will beep once the calibration is completed.