Information about DIN EN 62353 (VDE 0751-1:2008-8)
Testboy
®
TV 470
71
ENGLISH
Equivalent patient leakage current
Allowable values from long-term flow equiva-
lent patient leakage current
Value in mA
Application part
B
BF
CF
Equivalent patient leakage current.
–
5.0
0.05
Note 3
: Mobile X-ray machines with mineral insulation.
Important notes
|
Equivalent device leakage current insulation test is only valid when all
circuits in the device are activated.
|
The application part type determines the limit value.
If all safety-relevant parts are not reached with the equivalent device leakage
current, then a leakage current measurement should be made instead. Use
either the direct method or the indirect method
. With the indirect method,
the DUT has to be in an isolated position.