13
2.1 Checking the shipment
Upon receiving the shipment, first check the packaging and outer equipment cover for visible damage. Also,
check packaging and casings of peripherals (if you ordered any). Record in writing any defects which were
possibly caused in transit. If the shipment shows damage or is not complete, immediately advise the shipping
agency and/or your dealer.
2.2 Delivery contents
The standard delivery contains the NSG 5500 with connection cables and the instruction manual. Optional
modules are installed in the NSG 5500 if ordered. Possibly, there is also optional equipment, e.g. computer
or printer.
2.3 Operating position
The NSG 5500 must be placed in upright position firmly and securely during operation.
2.4 Line voltage connection and grounding
2 INSTALLATION
DANGER!
Connect the NSG 5500 only to line voltage conforming to the power specification
given on the type label.
If the grounding is interrupted inside or outside of the equipment, the NSG 5500
will become a source of danger. Additionally, the output of the device may not
conform with the relevant test standards when the grounding of the power plug
is interrupted. Carefully observe grounding precautions. The plug on the power
cord may be replaced by qualified personnel only.
The power input of the tester is located on the rear side of the NSG 5500 (see picture
3-4). The power input
is secured by two 3.15 A time-lag fuses (220–240 V range) or two 6.3 A time-lag fuses (110–120 V range). For
details, refer to chapter 4.3 Protection.
The power plug and outlet must have grounding contacts. When the NSG 5500 is brought from cold to
warm environment, the ensuing condensation may bring about dangerous conditions.
DANGER!
The NSG 5500 may only be switched on after all parts have acclimatized.