![Tescom TC-5915AP Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/tescom/tc-5915ap/tc-5915ap_user-manual_1090866032.webp)
24
Shield Box Operation with Fixture
4.2.4 PASS/FAIL LED Indicator
Green and red LED indicators will be on if the command "PASS" or "FAIL" is sent through RS-232C
while the lid is closed.
This function can be useful to visualize the result when the Shield Box is controlled via RS-232C.
The red LED will glow automatically if the lid close sensor does not detect the lid or a problem
occurs during the fixture operation.
4.3 Shield Box Operation with Fixture
The fixture operation mode of standard TC-5915AP is set as 'Handler Mode: Off', 'Probe Mode:
Timer' at the time of release. (You can check it with the query 'SOLENOID MODE?') You need
to configure the fixture operation mode if you are to use it.
4.3.1 Fixture Operation Modes
The handler and the probe of fixture can operate in 3 modes each, and the operation is automatic
according to the open or close of Shield Box.
The operation of fixture can be controlled with commands via RS-232C. Refer to the list of commands
and each command in detail at