14
4-4-2 Result modes
The bar graph display always shows each axis (X, Y or Z)
the instantaneous measured dynamic range value. The
digit display shows the instantaneous or result according to
one of three modes which can be selected :
•
Instantaneous: The display shows the last value measured
by the sensor, no symbol is displayed.
•
Maximum instantaneous (MAX): The digital display
shows the highest instantaneous value measured, the
“MAX” symbol is displayed.
•
Maximum average (MAX AVG): The digital display shows
the highest average value measured, the “MAX AVG” symbol
is displayed.
t
Max. Hold
Average
Actual
Max. Average
Field strength
4-4-3 Calibration factor (CAL)
The calibration factor CAL serves to calibrate the result
display. The field strength value measured internally is
multiplied by the value of CAL that has been entered and
the resulting value is displayed. The CAL setting range is
from 0.20 to 25.00.