3-2 Electrical specifications
Unless otherwise stated, the following specifications hold
under the following conditions:
•
The meter is located in the far-field of a source, the sensor
head is pointed towards the source.
•
Ambient temperature: +23
℃
±
3
℃
•
Relative air humidity: 25% to 75%
Sensor type: Electrical field (E)
Frequency range: 10MHz to 8GHz
Specified measurement range:
•
CW signal (f > 10MHz): 20mV/m to 108.0V/m,
53
µ
A/m to 286.4mA/m,
1
µ
W/m
2
to 30.93W/m
2
,
0
µ
W/cm
2
to 3.093mW/cm
2
Dynamic range: Typically 75dB
Absolute error at 1 V/m and 10 MHz:
±
1.0dB
Frequency response:
•
Sensor taking into account the typical CAL factor:
±
1.0dB (10MHz to 1.9GHz)
±
2.4dB (1.9GHz to 8GHz)
•
Isotropy deviation: Typically
±
1.0dB (f>10MHz)
•
Overload limit: 10.61mW/cm
2
(200V/m)
•
Thermal response (0 to 50
℃
):
±
0.5dB