“MIN” "T1" , "T2" , "T1-T2" , "

" , "

" ,low battery " 

 " etc. 

2).

 

ON/OFF:

 The on/off key will turn the thermometer on and off. 

3).

 

MAX/MIN:

 

The max/min key will turn the display into Maximum and Minimum 
mode. Press it once and the meter will show the Maximum reading. 
Press it again and the meter will show the Minimum reading. If we 
hold this key for 2 seconds, the meter will be back to the normal 
mode.  If MAX/MIN mode is selected, except HOLD and 

/

℉, 

all 

function key will be disabled. 

4).

 

HOLD:

 

Press this key will hold the data. If we press and hold this key under 
power off then press the on/off button, the thermocouple type will 
be set as J-type when it is on. The default power on thermocouple 
type is set as K-type. 

5).

 

T1

 Thermocouple Input Connector. 

6).

 

T1/T2/T1-T2   :

 

When we first power the meter on, T1 will be shown. If we press 
this key once, the meter will show T2.  
If we press it again, T1-T2 will be shown. 
If we press it again, the meter will show T1, T2, T1-T2 in circulation. 
If this key been pressed again, it will turn the meter to default 
reading ( that will be T1). 

7).

 

REL:

 

Press this key will turn the meter into relative mode. The reading 
will first show zero and then the reading change relate to the setting 
time will be shown. 
Press it again will show relative value. 

If hold this key 2 seconds the meter will be back to the normal 
mode. 

8).

 

/

:

 

Press this key will change the temperature scale between 

 and 

.  

9).

 

T2

 Thermocouple Input Connector. 

10).

 

ZERO: 

The  ZERO  controls  allow  you  adjust  offset  value  to 

optimize measurement accuracy for a particular thermocouple at a 
particular temperature. 

4.

 

RECALIBRATION PROCEDURE 

The  thermometer  should  be  calibrated  once  a  year  to  ensure  its 

accuracy  is  within  specifications.  The  required  equipment  is  listed 
below: 

 Type J:    0.0

 adjust VR4. 

 Type K:    0.0

 adjust VR1. 

 Type K:199.0

 adjust VR3. 

 Type K: 1000

 adjust VR2. 

WARNING

 

o avoid  electrical  shock, do  not  use  this  instrument  when 
voltages  exceeding 24VAC or 60VDC are present.  The probe 
tip is electrically connected to the output terminals. 

 

Содержание TES-1306

Страница 1: ... or service the meter Use the meter only as specified in this manual otherwise the protection provided by the meter may be impaired Environment conditions Altitude up to 2000 meters Relatively humidity 90 max Operation Ambient 0 40 Maintenance Clearing Repairs or servicing not covered in this manual should only be performed by qualified personnel Periodically wipe the case with a dry cloth Do not ...

Страница 2: ...uracy are of reading degree at 18 to 28 with relative humidity up to 80 For single thermocouple measurements Function Resolution Range Accuracy Type K Type J 0 1 200 50 200 50 0 5 rdg 1 0 50 0 50 0 0 2 rdg 1 0 0 200 0 150 0 1 rdg 0 8 1 200 1370 150 760 0 2 rdg 2 0 2 328 58 328 58 0 5 rdg 2 0 58 32 58 32 0 2 rdg 2 0 32 392 32 302 0 1 rdg 1 6 2 392 2498 302 1400 0 2 rdg 3 For T1 T2 Measurements accu...

Страница 3: ...he display Break or no input indication _ _ _ _ appears on the display Low Battery Indication The is displayed when the battery voltage drops below the operating voltage Power Requirement 9 Volt battery NEDA 1604 or JIS 006P or IEC6F22 Battery Life typical 100hours Alkaline Battery Dimensions 135 L x 72 W x 31 H mm 5 3 L x 2 8 W x 1 2 H inches Weight Approx 235g with battery Accessories Battery In...

Страница 4: ...in the meter will show T1 T2 T1 T2 in circulation If this key been pressed again it will turn the meter to default reading that will be T1 7 REL Press this key will turn the meter into relative mode The reading will first show zero and then the reading change relate to the setting time will be shown Press it again will show relative value If hold this key 2 seconds the meter will be back to the no...

Страница 5: ...ion probe 50 to 1000 58 to 1832 2 2 or 0 75 3 6 or 0 75 3 2φ 150 metal sheath 100 Compensating wire TP K03 Surface probe 50 to 750 58 to 1382 2 2 or 0 75 3 6 or 0 75 100 Compensating wire 12 5φ x 94 handle TP K01 Available for general condition especially for complex and any place hard to reach 8 TP K02 Available for temperature measurement of liquid gels or air TP K03 Available for flat or curved...

Страница 6: ...9 ...

Страница 7: ...TES ELECTRICAL ELECTRONIC CORP 7F No 31 Lane 513 Rui Guang Road Neihu Dist Taipei Taiwan R O C Tel 02 2799 3660 Fax 886 2 2799 5099 E Mail tes ms9 hinet net http www tes com tw June 2000 4 ...

Отзывы: