Perform a quality control with the FINETOUCH Control Solution
Peel off the film seal from the tip case and attach the
Test TIP to the blood glucose meter. Do not discard the
film seal (you will need it afterwards for carrying out the
control measurement).
Shake the bottle of the FINETOUCH control solution
lightly. Remove the screw cap and put one drop (with a
minimum diameter of approximately 3 mm) of the control
solution on the silver side of the film seal, peeled off from
the tip case.
Immediately after placing the droplet on the film seal,
suck up the control solution using the Test TIP. After
hearing the beep, remove the Test TIP from the control
solution and place the blood glucose meter on a flat
surface.
When the measurement value is displayed, compare it
with the tolerance range indicated on the label of the
control solution. The measurement result must be within
this tolerance.