Bionova® MiniBio
Page 6
Instructions for use
Record Table
Use the following table to record firmware updates.
NOTE:
manufacture firmware is printed on a label found
at the bottom of the incubator and on its packaging.
Please, register your manufactory firmware in this manual
for future reference:
Manufacture Firmware:
Fecha
Versión
Operador
Start-up
1-Place the auto-reader on a firm surface, free from vibra-
tions, away from sunlight, currents of hot or cold air, che-
mical and corrosive or flammable substances. Do not pla-
ce the equipment in a way that disconnection of the plug
from the power supply could be difficult. Leave a gap of
at least 10 cm from the wall. Do not move the auto-reader
periodically or during its use. Connect the auto-reader to
a secure and stable electrical power source.
Do not wet or heat the auto-reader. If liquid is spi-
lled on the incubator, disconnect it and dry it im-
mediately. Before turning on, verify that all reading
positions of the MiniBio incubator are empty.
2-Turn on the incubator by connecting the end of the AC
of the power supply to the power source and then con-
nect the other end of the source (plug) on the rear of the
Bionova® MiniBio unit. The equipment will start by run-
ning the last time and temperature set-up.
3-Synchronize the Bionova® MiniBio incubator with local
time, using Bionova® software available in:
http://www.
terragene.com.ar/productos/infection-control/incuba-
tors/software-downloads/?lang=en
. Connect incubator
USB port to the computer and start the program. Biono-
va® MiniBio incubator unit will be synchronized with the
time and time zone in the PC. Synchronization can be done
as long as there is not an ongoing reading. See other con-
figuration methods in the
Time setting mode
section.
All devices are manufactured with the following
settings:
- Time zone: UTC +0:00
- Printing language: English
If you want to change the time zone, refer to
Modification
of time zone
section. If you want to change the print lan-
guage, refer to
Modification of print language
section.
4- Check that required incubation temperature is correct.
To change it, press button
for 3 seconds. Lights
corresponding to 37 ºC and 60 ºC will start blinking, indi-
cating that the temperature program must be selected.
Press the button
to select the incubation temperatu-
re program, 37 ºC or 60 ºC. After 3 seconds, temperature
modification will be accepted.
For 37 ºC, Bionova® MiniBio Auto-reader will automatica-
lly select 4 hs. final readout for all positions.
For 60 ºC the incubator allows choosing between
Ultra
(20 minutes), 30 minutes, 1 hour, 2 hours or 3 hours final
Содержание BIONOVA MiniBio
Страница 1: ...Rev 5 September 2018 MiniBio Innovation in technologies for Infection Control ...
Страница 2: ......
Страница 14: ......
Страница 15: ...Rev 5 Septiembre 2018 MiniBio Innovación en tecnologías para el Control de Infecciones ...
Страница 27: ......
Страница 28: ......
Страница 29: ...Rev 5 Setembro 2018 MiniBio Inovação em tecnologias para Controle de Infecção ...
Страница 41: ......
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ......