background image

 

 

RSR ECM

 

Fan/Filter Unit

       

Installation and Operation Manual    

Doc. 1788-25

 

© 

Copyright 2018 Terra Universal Inc.   

All rights reserved 

Revised November 2019 

 

 

Terra Universal, Inc.  •  TerraUniversal.com  •  800 S. Raymond Ave.  •  Fullerton, CA  92831  •  TEL: (714) 578-6000  •  FAX: (714) 578-6020 

4

 

7.0

 

Test Probe Terminals 

 

The test probe terminals are provided on the EC Motor Control Card to measure the 

motor RPM or check for a PWM signal using a multimeter. 

 

Manual or Analog Control Modes: 

To measure RPM, set the Address DIP Switches at a value greater than 1.  The test 

probe jacks will output 0-2000 mVDC to represent motor RPM. 

 

Example: 500 mVDC = 500 RPM 

 

To measure demand signal, change the Address DIP Switches to the value of 0. 

 

NOTE: Changing the address will not interrupt power to the Control Card. 

 

Network Control Mode: 

In the MODBUS® network control mode, the test probe terminals will always output  

0-2000 mVDC to represent motor RPM. 

 

8.0

 

120V vs. 220V Configuration 

 

 

120V Configuration 

1.

 

Black

 and 

Red

 wires are connected to the lower terminal. 

2.

 

Yellow Jumper Cable is installed. 

 

 

 

 

220V Configuration 

1.

 

Black 

and

 Red

 wires connect to the upper terminal as shown. 

2.

 

Yellow Jumper Cable is removed. 

 

 

 

 

9.0

 

Control Card Electrical Specifications 

 

Electrical Ratings 

Input Voltage, AC, 50/60Hz 

12 – 28V

1

 AC 

Input Voltage, DC 

14 – 30V DC 

Operating Current, AC 

35 mA AC 

Operating Current, DC 

30 mA DC 

Accessory Current 5V, 10V 

50

3

 mA DC 

Accessory Current V.RFU

100

3

 mA DC 

Refers to 

actual AC voltage values

 

V.RFU is rectified and capacitive-filtered, but is unregulated 

Maximum current values stated in support of a single connected load

 

 

“Stand Alone Operation” Power Source Ratings 

Supply Transformer Voltage Rating, 

12 – 24V AC 

Supply Transformer Power Rating 

2 VA 

Regulated DC Supply Voltage Rating 

14 – 30V DC 

Regulated DC Supply Power Rating 

2 W 

 

Extended Accessory Output Currents for Stated Input Voltage Condition 

Input Voltage 

Accessory Current, DC mA

Suggested 

AC/DC Source 

Power Rating 

5V 

10V 

V.RFU 

VA 

14V DC Regulated

 

300

 

300

 

500

 

20,20,30

 

30V DC Regulated

 

50

 

50

 

500

 

10,10,60

 

12V AC

1

 Transformer

 

100

 

100

 

100

 

5,5,5

 

28V AC

1

 Transformer

 

50

 

50

 

200

 

10,10,25

 

1

 

AC voltage spec refers to 

actual AC voltage values

, i.e., not transformer ratings, etc.

 

2

 

Current values stated in support of a single connected load.

 

 

10.0

 

EC Motor Specifications 

 

Specifications 

Horsepower 

1/3 HP 

Voltage 

120/240V, 50/60Hz 

Speeds 

300 – 1250 RPM 

Inputs 

2 Way Serial Communication & PWM 

Frame

 

NEMA® 48 

Enclosure 

Continuous Air Over 

Mounting 

Belly Band 

Amperage 

4.2/2.7 Amps 

 

 

 

 

Содержание RSR ECM

Страница 1: ... configurations with other products systems such as ductwork or hoods 2 0 Operation The unit consists of a 1 3 HP electronically commutative motor driving a forward curved centrifugal impeller Air is drawn into the unit through a pleated pre filter and in turn to an insulated chamber where it is then directed to an exit HEPA filter with a built in diffuser The unit can also be equipped with a 3 8 ...

Страница 2: ...he screen to the fan module housing B Next rotate the metal tabs that help hold the HEPA filter in place see Photo 3 Photo 3 Filter face grille s metal tabs C Gently let the used filter drop down and set aside The filter assembly is surrounded by a channel containing gel seal Wipe any residual gel from the knife edge of the fan module still in the ceiling to prepare for the replacement filter Posi...

Страница 3: ...nimum 60 RPM measured Industry Standard MODBUS Networking RTU Protocol RS485 9600 8 n 1 6 2 Control Modes The EC Motor Control Card offers three different control modes for operating the FFU MODBUS RTU networking 0 10 VDC analog control or manual control through an onboard potentiometer Manual Control Mode The onboard potentiometer can be manually adjusted at the EC Motor Control Panel using a Phi...

Страница 4: ...ble is removed 9 0 Control Card Electrical Specifications Electrical Ratings Input Voltage AC 50 60Hz 12 28V1 AC Input Voltage DC 14 30V DC Operating Current AC 35 mA AC Operating Current DC 30 mA DC Accessory Current 5V 10V 503 mA DC Accessory Current V RFU2 1003 mA DC 1 Refers to actual AC voltage values 2 V RFU is rectified and capacitive filtered but is unregulated 3 Maximum current values sta...

Страница 5: ...separately For Terra s Policy and Procedures for Returning Goods see Terra s Internet site www TerraUniversal com Generally customers can improve the chance of collecting on a freight claim by following these procedures 1 formally requesting that the carrier inspect the shipment immediately upon suspecting damage or shortage to verify condition 2 notifying the carrier upon discovery of concealed d...

Отзывы: