Quick-Start Operating Guide
Series 310 Glove Box
©
Copyright 2017
Terra Universal Inc. All rights reserved. •
Revised Feb. 2017
• Document No. 1788-10
v2.0tm
Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020
4
These set-up procedures apply to most common applications.
Uncrating and Setting up the Glove Box
Make sure that your glove box is placed on a level, stable surface, away from heat or chemicals that could damage it.
For Use with Remote Ventilation Filter/Blower
In this application, the glove box is shipped with a conduit flange to allow easy conduit connection.
1.
Close the vacuum valves located on the top of the glove box.
2.
Connect exhaust conduit to the conduit flange. The standard flange is for 6"-diameter hosing.
3.
Connect the other end of the conduit to Terra's remote filter/blower module. The filter/blower module should be
equipped with a purification (activated charcoal) filter appropriate to the exhaust fumes in your application.
To prevent dangerously low oxygen levels and risk of asphyxiation, nitrogen-purged systems should only
be installed in a well-ventilated area.
WARNING
To in-house exhaust system
Ducted Exhaust Fume Ventilation
For safe in-door release
Ductless Exhaust Fume Ventilation