Quick-Start Operating Guide
Series 310 Glove Box
©
Copyright 2017
Terra Universal Inc. All rights reserved. •
Revised Feb. 2017
• Document No. 1788-10
v2.0tm
Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020
3
For automated vacuum control, select the Vacuum Control
Module and Vacuum Pump Connection Kit. For automated
backfill and purging, select the Purge Control Connection
Kit, Dual Purge System, and Trace Oxygen Analyzer, which
work in tandem to provide oxygen levels down to 1 ppm.
Right- or left-side air locks allow transfer of materials into
the glovebox without breaching the controlled environment.
Dual gaskets on both the tilt-up window and air lock doors
provide a safe, positive seal while allowing full access to the
internal chamber to insert or remove materials or
equipment. The side doors open inward to allow an
operator to move items in and out easily.
Changeable Sleeve and Glove Combination
The front panel of these glove boxes features 10" flanged arm ports that allow easy connection of Terra’s Changeable
Sleeve and Glove Combination.
Air Locks
These units can be mounted on either side of the chamber to provide a safe, practical way of moving materials into
and out of the work area. Materials are loaded into the air lock by means of an external door; an operator then opens
an internal door inside the glove box to bring materials into the work area. The standard air lock features its own gas
inlet, system pressure, and an Automatic RB Valve port in case your application requires air lock purging.
Tilt-Up Viewing Windows
The Series 310 Glove Box comes standard with a Tilt-Up Viewing Window that swings open to allow easy access.
3.0
Installation
Terra Glove Boxes are normally delivered with all
specified components as a complete turnkey system.
In this case, you need to only make the required
vacuum and power connections.
We recommend, however, that you review all of the
information in this section. It provides guidelines that
will prove helpful if you ever add to your system or if
you need to disassemble system modules in order to
perform maintenance procedures or move large
equipment into the glove box.
Carefully uncrate the glove box and accessories and
visually inspect for damage, both inside and outside.
Be sure not to throw away any important items with
the packing material. Any damages should be
reported according to the procedures outlined in the
shipping agreement.
Safety Relief Valve
Vacuum
Valves
Gloves Removed,
Iris ports, or blank openings