25
8.5. Capacitance measurement
The capacitance value is obtained by measuring the insulation resistance.
After finishing measuring (When the
key has been pressed), wait the
high voltage LED turn off, and press the
key and the capacitance
value will be exhibited on the display.
Voltage
Capacitance
500 V
50 nF up to 10 µF
1,000 V
50 nF up to 5 µF
2,500 V
30 nF up to 2 µF
5,000 V
30 nF up to 1 µF
Precision: ± 10% of the measured value ± 3 digits
Note
: “
0
” will be exhibited on the display when measuring values lower
than 50 nF
8.6. Hold
This function allows holding the last performed reading on the display at
the moment when pressing the
key, without interrupting the test.
When this key is pressed again, the equipment updates the resistance
and chronometer values. The led of
key and the letter
H
on the
display indicate that the function has been activated