21
Example:
The display exhibits the test number (12), the measured resistance value
(602
GΩ
)
, the voltage selected of 2500 V, the elapsed time (6:25
minutes), the applied voltage (2.51 kV), the leakage current (4.17 nA),
date and time.
NOTE:
If the resistance measured is greater than 5 T
Ω
@ 5 kV, the
following message will be exhibited:
R > 5T
Ω
ATTENTION:
Please, never connect or disconnect the test leads with the
equipment under operation or while the High Voltage led is on. If there is
a need to modify the connections, this should be done with the equipment
disconnected and with discharged potentials (High Voltage led off)