Calibrate Instrument 

*IMPORTANT:

  

T

HIS FEATURE SHOULD ONLY BE USED BY A CALIBRATION FACILITY USING CERTIFIED AND 

TRACEABLE DEAD WEIGHTS

.

  

D

O NOT USE THIS FEATURE FOR A SIMPLE ACCURACY CHECK

NOTE 1: TX-1 instruments are calibrated for Wire and Filaments through a 
factory process not available to users. Use the calibration procedure below 
to add or change a calibration in the CUSTOM menu. 

NOTE 2: When calibrating your TX-1 instrument, tension values must be 
entered in grams.   

1.  To add or change a calibration in the CUSTOM menu of your TX-1 meter, 

first hang the material for the calibration as described in “General Calibration 
Precautions” above. (You will suspend the weights later in the process.) 

2.  Access the setup menu: Turn on your meter and press the DOWN ARROW 

until SETUP is highlighted, then press ENTER. 

3. 

If there isn’t yet a CUSTOM NAME in the menu for the material, then add one 
as follows: 

o

 

Scroll down until CUSTOM NAMES is highlighted, and press ENTER. 

o

 

Scroll down to CUSTOM1 (or to the next available CUSTOM entry), 
and press ENTER. 

o

 

Using a combination of the UP, DOWN and ENTER buttons, key in a 
name for your custom material. 

4.  Add or change the calibration for your CUSTOM material: 

o

 

Press the ESCAPE button until only SETUP is showing, and press 
ENTER. 

o

 

Scroll down to RE-CAL TENSION, and press ENTER 

o

 

Press the ENTER, UP, ENTER buttons. (This is the password.) 

o

 

Scroll up or down to select the CUSTOM NAME for the material want 
to calibrate, and press ENTER. 

o

 

Enter the number of calibration points you will be using (at least five, 
and no more than ten). Zero counts as the first point. 

o

 

ENTER WEIGHT for each data point:  

 

Zero is the first value and is already programmed.  

 

Each successive data point should be a higher weight than 
the previous one.  

 

Data points are typically distributed evenly along the tension 
range. (Example: With 5 data points for 1000 grams, use 0, 
200, 400, 600 and 800.) 

o

 

When the display reads Zero Unit, hold the instrument (NOT engaged 
on the material) in the attitude that you plan to do the calibration (likely 
horizontal). Then, press ZERO. 

o

 

As the display prompts you, suspend the correct weight from your 
material. Then clamp the instrument on the material and press 
ENTER. 

o

 

Repeat for each weight as prompted. 

Содержание TX-1000-1

Страница 1: ...TX 1 Series Instrument www tensitron com pg 1 Operating Instructions TX 1 Series Digital Wire Filament Tension Meter...

Страница 2: ...ing Instrument onto material 9 6 DISPLAY SCREEN 10 SETUP Screen 10 o LCD Refresh Rate 10 o Data Averaging 10 o Cal Analog Optional feature 10 o Cal Length Optional feature 10 o Custom Names 10 o Resol...

Страница 3: ...assistance please call us at 303 702 1980 or find us online at www tensitron com 1 WARRANTY POLICY STANDARD EQUIPMENT WARRANTY Tensitron warrants that all Tensitron manufactured equipment will be fre...

Страница 4: ...proximately 12 hours depending on backlighting intensity o Operating Temperature 32 F to 120 F If these values are exceeded battery charge discharge rates will decline However this could be offset by...

Страница 5: ...nd any other condition that may affect operation Do not use a damaged instrument Tag damaged instrument DO NOT USE until repaired For repair send instruments directly to Tensitron Inc o Guard against...

Страница 6: ...s Newtons or cN cN only available on TX 125 1 series instruments o Instrument includes calibrations for numerous wires and filaments o Extremely accurate custom calibration using up to a 10 point line...

Страница 7: ...ly available from stock o Numerous application specific models available o Durable lightweight carrying case with protective foam inserts See Figure 2 o All instruments are factory calibrated and read...

Страница 8: ...power the instrument off c For additional information on logging and viewing data refer to Sec 6 Display Screens Data Logging 4 Read Tension Variations in materials affect tension readings It is essen...

Страница 9: ...you engage it to the tensioned material Then press ZERO Engaging Instrument onto material WARNING DO NOT EXCEED THE MAXIMUM TENSION RANGE OF THE INSTRUMENT OR DAMAGE WILL OCCUR Engage the Instrument...

Страница 10: ...ote This feature is primarily used in processes where the material to be checked is constantly moving such as coil winding and textiles o Cal Analog Optional feature Use this feature to calibrate the...

Страница 11: ...its and press ENTER Select your preference Meters or Feet and press ENTER o Re Cal Tension See description at the end of this booklet ONLY PERFORM THIS IF YOU ARE A QUALIFIED CALIBRATION HOUSE USING N...

Страница 12: ...t tension readings are logged Scroll to DATA LOGGING and press ENTER Scroll to DURATION and press ENTER Use the Up and Down arrows to change the Duration that you want data stored and press ENTER Mini...

Страница 13: ...measured for tension AWG Inches or mm o Select Material Filaments Units This feature allows the operator to select the unit for describing diameter of filaments to be measured for tension Inches or mm...

Страница 14: ...minimum repeatable accuracy standards Check Accuracy 1 To check the accuracy of your meter first hang the test material and suspend weights from it as described in General Calibration Precautions abo...

Страница 15: ...he next available CUSTOM entry and press ENTER o Using a combination of the UP DOWN and ENTER buttons key in a name for your custom material 4 Add or change the calibration for your CUSTOM material o...

Страница 16: ...733 S Bowen Street Longmont CO 80501 USA Phone 303 702 1980 Fax 303 702 1982 E mail sales tensitron com Web Site www tensitron com...

Отзывы: