TensCare Sure PRO Скачать руководство пользователя страница 28

27 

 

is  displayed,  and  the  unit  returns  to 
waiting mode. 

To  view  the  memory,  press 

+Menu

  or      

-Menu

 and cycle through parameters to 

the screen with 

MONTH

 showing in the 

middle  and  the  warning  triangle 

  at 

bottom centre. 

Use  the 

+

  and 

-

    buttons  to  cycle 

through recordings.  

To clear the memory, press and hold the 

+Menu

 and 

-Menu

 buttons together for 

3 to 5 seconds while in memory mode. 

 

PAUSE 

Pressing  the  ►||  button  while  a 
programme  is  in  use  stops  the 
stimulation  and  the  timer.  The  symbol 
►|| is displayed. Press again to resume 
the programme. Stimulation will re-start 
at  75%  of  previous  intensity  setting.  If 
the  device  is  left  in  PAUSE  for  more 
than  15  minutes,  it  automatically 
switches off. 

 

OPEN CIRCUIT DETECTOR  

Sure PRO 

controls the connections and 

the  contact  between  the  probe  or 
electrodes and your body. 

This  is  to  avoid  sudden  changes  in 
stimulation if a broken connection is re-
established. 

If the lead or probe does not have good 
contact  or  become  detached  and  the 
intensity  is  set  to  greater  than  10mA, 
the  warning  triangle  symbol 

  will 

flash, the zeroes in the display will flash, 
and the intensity will automatically reset 
to zero. 

Check the lead and try using a water-
based lubricating gel such as TensCare 
Go Gel (

K-GO

) with the probe. 

If the problem persists, see section 20.  

 

To preserve battery life, the 

Sure PRO

 

automatically switches off if left at zero 
intensity for more than 2 minutes. 

 

LOW BATTERY  

A  low  battery  warning  symbol 
will display when  you need to 
recharge the battery.  

 

Automatic Keypad Lock 

If  no  button  used  for  10 
seconds  and  intensity  is  set 
above  zero  the  keypad  will 

lock 

automatically. 

This 

prevents 

accidental changes to settings.

 

Press 

Ch1

▼ or 

Ch2

▼button to unlock. 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание Sure PRO

Страница 1: ...0 INSTRUCTIONS FOR USE READ CAREFULLY BEFORE USE...

Страница 2: ...probe Insert the vaginal probe Press the ON button to switch the device on Regulate the output intensity with the buttons and Press the OFF button to switch the device off ON Button P UP DOWN OFF Sele...

Страница 3: ...tested products for the applications in the areas of gentle electrotherapy muscle toning continence management and pain relief during labour Please read these instructions for use carefully and keep...

Страница 4: ...IMULATION 20 10 4 MANUAL PROGRAMMES 21 11 CONTENT 23 12 UNIT INFORMATION 24 12 1 CONTROLS DISPLAY 24 12 2 OPERATING INSTRUCTIONS 25 13 SETTING UP AND USING THE SURE PRO 28 13 1 CHARGING THE BATTERY 28...

Страница 5: ...ue number so that the device can be identified Atmospheric Pressure indicates the atmospheric limits to which the medical device can be safely exposed Date of Manufacture indicates the date which the...

Страница 6: ...continence including stress urge and mixed types as well as post prostatectomy urinary incontinence in men Additionally it may help improve sexual intimacy by toning the pelvic floor muscles Chronic p...

Страница 7: ...of a bowl in shape They run from the pubic bone in the front to the end of the spinal column or tail bone in the back The pelvic floor muscles Assist in supporting the abdominal and pelvic organs Work...

Страница 8: ...n stop accidental urine faeces or wind leakage It can also reduce your risk of prolapse improve your recovery from childbirth and gynaecological surgery and increase your sexual pleasure A continence...

Страница 9: ...ing your Sure PRO pelvic floor stimulator in conjunction with Kegel exercises will give you a better understanding of how they work and how to get the greatest benefit from them 5 TYPES OF INCONTINENC...

Страница 10: ...natural pain relief mechanism to reduce pain without side effects 6 HOW EMS WORKS E M S stands for Electrical Muscle Stimulation and has successfully been used in medical rehabilitation and training i...

Страница 11: ...0 50 Hz can put a high level of strain on the muscle thus promoting the muscular structure Each preset programme has a specific frequency and pulse width that will offer the best results for the type...

Страница 12: ...iod pain and endometriosis see Ova cystitis sports injuries fibromyalgia and neuralgia plantar fasciitis post operative pain TMJ disorder diabetic neuropathy osteo arthritis and sometimes non painful...

Страница 13: ...ur healthcare professional before use Do NOT use when driving operating machinery or similar actions needing muscular control Loose electrode pads damaged leads or sudden changes in contact may cause...

Страница 14: ...intra vaginal or intra anal use Improper treatment or cross infection may occur Caution It is important that the vaginal probe is cleaned after each use Ineffective cleaning may lead to irritation or...

Страница 15: ...ads other than those recommended by the manufacturer Performance may vary from specification Electrodes with smaller surface area may cause tissue irritation TO KEEP YOUR DEVICE IN GOOD WORKING ORDER...

Страница 16: ...dium frequencies 20 50 Hz can put a high level of strain on the muscle thus promoting the muscular structure Endorphin release programme PAIN only A low frequency of 4 or 10 Hz allows for the release...

Страница 17: ...00 1 10 1 20 0 30 5 60 15 CSTM 3 4 90 50 400 1 10 1 20 0 30 5 60 PFW Frequency Hz Pulse Width s Work sec Rest sec Default duration min A 20 250 s fixed 4 6 2 20 250 s stepped 4 6 5 3 250 s fixed 4 4 5...

Страница 18: ...1 mA 2 STRE 2 Stress Incontinence aftercare Having restored your pelvic floor muscles to an excellent condition with STRE 1 you will want to keep them toned and strong Regular use of this programme a...

Страница 19: ...trol but experience a lack of sensitivity This can be due to nerve damage and is a problem that may be helped with the programme The sensation is like a strong drawing in of the muscles of the vagina...

Страница 20: ...aecal incontinence Uses 10Hz See Section 10 3 for more information about this programme 12 TIBN 2 Transcutaneous tibial nerve stimulation Same as 11 but at 20 Hz For the programmes that cause muscle c...

Страница 21: ...th can be increased during the course of the treatment as you become accustomed to the sensation The powerful muscle contractions caused by electrical stimulation sometimes give rise to training aches...

Страница 22: ...priate settings could cause discomfort or muscle injury or aggravate Urge incontinence The programmes 13 14 and 15 respectively CSTM 1 CSTM 2 CSTM 3 are the manual programmes The effect of electrical...

Страница 23: ...in thickness as well as quality of electrode contact The actual setting should be effective but should never produce any unpleasant sensation such as pain at the site of application In pain relief and...

Страница 24: ...e fibres can only activate a limited number of times a minute about 1000 without becoming fatigued If the warning triangle is displayed at any other time see the troubleshooting in section 20 11 CONTE...

Страница 25: ...parameter Keypad lock Low battery Intensity Ch1 Intensity Ch1 Programme selector Menu parameter adjust Pause Programme Pulse width Time setting Pause Intensity Ch2 Warning ON OFF Intensity Ch2 Menu p...

Страница 26: ...rease strength in steps of 1 mA press and release the top part of the two channel buttons marked Ch1 and Ch2 To decrease the strength press and release the lower part of the buttons marked Ch1 and Ch2...

Страница 27: ...ers The selected parameter will flash Press or to adjust the setting Press to accept the change and the flashing will stop Parameters cycle through Hz s WORK REST RAMP TIMER DATE TIME MANUAL PROGRAMME...

Страница 28: ...ontact between the probe or electrodes and your body This is to avoid sudden changes in stimulation if a broken connection is re established If the lead or probe does not have good contact or become d...

Страница 29: ...appear on the charging cradle that means the battery is being charged When the battery is charged the indicator light on the cradle will change from red to green Storage Remove the battery from your S...

Страница 30: ...3 CONNECTING THE PROBE 1 Decide whether to use one channel or two Unless you are told otherwise by your medical advisor you will only want to use one Some treatment protocols use both vaginal and anal...

Страница 31: ...ge Note The flange should not be inserted into the vagina and should remain outside of the vagina at all times 13 5 TRAINING SESSION 1 Press and hold the ON OFF button on the control unit for at least...

Страница 32: ...your fatigue resistance Be careful not to overdrive the muscles this could cause painful aches the next day Note If you have aching or cramping muscles stop treatment until the symptoms go away then r...

Страница 33: ...aim to progress towards graduated active exercises in order to improve pelvic floor muscles strength and endurance and to regain function You may benefit from the Sure PRO if you either have no active...

Страница 34: ...rrect contact during the programme For Faecal incontinence the aim is to stimulate the external sphincter and or pubo rectal muscle so circular electrodes should be placed so that the external ring is...

Страница 35: ...SSORIES Expected Service Life The machine will often last for more than 5 years but is warrantied for 2 years Accessories lead wire probe and battery are not covered by the warranty Lead life depends...

Страница 36: ...defects for 2 years from date of purchase This warranty is void if the product is modified or altered is subject to misuse or abuse damaged in transit lack of responsible care is dropped if incorrect...

Страница 37: ...chase receipt Please ensure all relevant details are completed before sending your unit in for service Please ensure your contact details are still current and include a brief description of the probl...

Страница 38: ...tensity will return to 0 mA and the warning triangle will flash This safety feature will prevent the machine from giving any uncomfortable stimulation should the contact break between the machine and...

Страница 39: ...you can feel it The intensity increases in very small steps of 1 mA ii You may have reduced sensitivity due to previously damaged or desensitised pudendal nerves this can happen in childbirth or some...

Страница 40: ...n Internal power source Environmental Specifications Operating Storage Temperature range 5 to 40 C Humidity 15 to 93 RH Atmospheric Pressure 700 to 1060 hPa Temperature range 25 to 70 C Humidity Up to...

Страница 41: ...ANTY NAME ADDRESS POSTCODE DAYTIME TELEPHONE E MAIL MODEL DATE OF PURCHASE ATTACH PROOF OF PURCHASE DO NOT SEND IN PROBE OR ELECTRODE PADS RETAILERS NAME RETAILERS ADDRESS RETAILERS POSTCODE BRIEF DES...

Страница 42: ...41 NOTES...

Страница 43: ...e limited https plus google com TenscareLtdEpsom https uk pinterest com TensCareLtd https tenscareltd wordpress com https www youtube com channel UCzpik9dmLlJ3j0aHOpQ 0sg EC Declaration of Conformity...

Страница 44: ...43 Distributed by Manufactured by TensCare Ltd 9 Blenheim Road Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 www tenscare co uk Pub No I SP2 UK Version 3 0 06 18...

Отзывы: