29
Sudden
change in
sensation
If you disconnect and re-
connect a few minutes later,
the signal will feel quite a lot
stronger.
Always return strength to zero after
disconnecting the lead or the probe.
The patient is an intended operator. There are no user-serviceable parts inside the
unit, and no calibration is required.
If the above review has failed to resolve your problem, or to report unexpected
operation or events, or to provide feedback, call TensCare or your local dealer
(address on back cover) for advice. Contact TensCare customer service on +44 (0)
1372 723 434. Our staff are trained to assist you with most issues you may have
experienced, without the need to send your product in for service.
European Medical Device Regulation requires that any serious incident that has
occurred in relation to this device should be reported to the manufacturer and the
competent authority in your country. This can be found at:
https://ec.europa.eu/docsroom/documents/36683/attachments/1/translations/en/rendi
20. GENERAL SPECIFICATION
Waveform
Asymmetrical rectangular
Amplitude
(over 500 Ohm load)
99.0 mA
+/- 10%
Max intensity
50V zero to peak across 500
Ω
. Setting 0-99.0 mA in steps
of 00.5 mA
Constant voltage over 470-2000 Ohm
Constant current over 160-470 Ohm
OC cutout below 160 Ohm.
Max pulse energy
Total output limited to 25
μ
C per pulse
Output plug
Fully shielded: touch proof mini USB
Channels
Single channel
Battery
Rechargeable Li-ion battery
Model: BL-4B
Manufacturer: YJ Power Group Limited
Ratings: 600 mAh, 2.22Wh, 3.7V
Battery life
At least 15 hours at 50 mA, 300
μs
, 50Hz
Weight
90g
Dimensions
100 x 65 x 18 mm
Safety Classification
Internal power source. Designed for continuous use.
Environmental Specifications: