14
medical device that led or might have
led to the death of a patient, serious
deterioration of state of health of a
patient, or a serious public health
threat, should be reported immediately
to the manufacturer, and to the
Competent Authority in the Member
State in which the user resides.
8. INFORMATION ABOUT
THE PROGRAMME
SETTINGS
Each programme has its own
combination of Frequency and Pulse
Width settings which allow for different
sensations through the probe or
optional electrode pads and help
treating the different types of
incontinence.
•
Frequency (measured in Hz -
pulses per second)
Low frequencies (1-10 Hz) have a
purifying and relaxing effect through
individual contractions.
Medium frequencies (20-50 Hz) can put
a high level of strain on the muscle, thus
promoting the muscular structure.
•
Pulse Width (measured in
μs
-
millionths of a second)
The
Elise 2
unit has pulse widths of 200
to 300
μs
. Generally speaking, the
higher the pulse width, the more
"aggressive" the stimulation feels, if the
pulse width is set high enough, it will
usually elicit a muscle contraction,
which is required for an effective toning
of the pelvic floor muscles.
9. PROGRAMMES
9.1 PROGRAMME SETTINGS
9.2. PRESET PROGRAMMES
The
Elise 2
has four pre-set
programmes. One for each type of
incontinence (stress, urge and mixed),
and one for toning the pelvic floor
muscles (
TONE
).
STRESS INCONTINENCE:
Shown on the screen as:
STRES
The
STRES
incontinence programme
strengthens the muscles of the pelvic
floor using gentle stimulation. Once
these muscles are stronger, they are
better able to resist urinary leakage
caused by external pressure being
applied to the bladder such as with a
cough, sneeze or physical exertion.
The stimulation causes the muscles to
contract and work. This builds their
strength. Like other fitness training,
successful treatment requires
stimulation once a day for one to three
Prog
STRES URGE MIXED TONE
Freq.
(Hz)
50
10
10/
50
35
Pulse
width
(
μs
)
300
200 200/
300
250
Ramp Up
& Down
(s)
1
C
ons
tant
Con/
1
2
Plateau
(s)
5
Con/
5
3
Rest (s)
10
Con/
10
6
Default
duration
(min)
20
C
ont
i-
nuous
10/
10
20
Содержание Elise 2
Страница 1: ...0 INSTRUCTIONS FOR USE READ CAREFULLY BEFORE USE...
Страница 130: ...129 NOTES...