11
1.
Insert the red test lead into the
HzV
ȍ
terminal and
the black test lead into the
COM
terminal.
2.
Set the rotary switch to
ȍ
, resistance
measurement (
ȍ
) is default (or press
SELECT
button to select
ȍ
measurement mode)
3. Connect the test leads with the object being
measured.
The measured value shows on the display.
Note
z
The test leads can add 0.1
ȍ
to 0.2
ȍ
of error to
resistance measurement. To obtain precision
readings in low-resistance measurement, that is
the range of 400.0
ȍ
, short-circuit the input
terminals beforehand, using the relative value
function button
RESET
to automatically
subtract the value measured when the testing
leads are short-circuited from the reading.
z
For high-resistance measurement (>1M
ȍ
), it is
normal taking several seconds to obtain a stable
reading.
z
If
ȍ
reading with shorted test leads is not
0.5
ȍ
,
check for loose test leads, incorrect function
selection, or enabled Data Hold function.
z
The
LCD
displays
OL
indicating open-circuit for
the tested resistor or the resistor value is higher
than the maximum range of the Meter.
z
When resistance measurement has been
completed, disconnect the connection between
the test leads and the circuit under test.
Testing Diodes
(see figure 5)
Use the diode test to check diodes, transistors, and