When measuring at 20
Ω
and 200
Ω
range, the test clips
can add 0.1 to 0.3
Ω
error to resistance. To obtain precise
readings in low-resistance measurements, short circuit
the input terminals beforehand and record the reading
obtained (called this reading as X). (X) is the additional
resistance from the test clips.
Then use the equation:
measured resistance value (Y) – (X) = precision readings
of resistance.
The Meter displays
“1”
when there is no input, for
example, open circuit situation.
For high resistance measurement (>1M
Ω
), it normally
requires several seconds to obtain a stable reading.
When resistance measurement has been completed,
disconnect the connection between the test clips and
the circuit under test and remove the test clips from the
input terminals of the Meter.
l
l
l
12
Note
Model 72-8150: OPERATING MANUAL