8
TECHNICAL INDEX
•
Accuracy: ± (a% r b count).
•
Environment temperature: 23°C ± 5°C (73.4°F ± 9°F), relative humidity: ≤75%.
DC VOLTAGE MEASUREMENT
Range Resolution Accuracy
200.0mV
0.1mV
± (0.7% + 5)
2.000V
1mV
± (0.7% + 3)
20.00V
10mV
200.0V
100mV
600V
1V
•
Input impedance: approx. 10MΩ.
• As the input impedance on the meter is high,
when measuring a source of lower voltage
and lower impedance (less than 10MΩ), the
measurement may be unstable for a short
time, but will stabilize.
• Warning: Maximum input voltage: ±600V
AC VOLTAGE MEASUREMENT
Range Resolution Accuracy
2.000V
1mV
± (1.0% + 3)
20.00V
10mV
200.0V
100mV
± (1.0% + 3)
VFC mode:
± (4.0% + 3)
600V
1V
± (1.2% + 3)
VFC mode:
± (4.0% + 3)
•
Input impedance: approx. 10MΩ.
• Show true virtual value: frequency response:
45~400Hz.
• Accuracy guarantee range: 5-100% range,
short circuit allows <10 residue readings.
• Non-sinusoidal wave counts add error by
crest factor:
- When crest factor is 1~2, add 3%.
- When crest factor is 2~2.5, add 5%.
- When crest factor is 2.5~3, add 7%.
• Warning: Maximum input voltage: 600Vrms.
RESISTANCE MEASUREMENT
Range
Resolution Accuracy
200.0Ω
0.1Ω
± (1.0% + 2)
2.000kΩ
1Ω
20.00kΩ
10Ω
200.0kΩ
100Ω
2.000MΩ
1kΩ
± (1.2% + 3)
20.00MΩ
10kΩ
• Range: measured value = measurement
display value - meter pen short circuit value.
• Open circuit voltage is about 1V.
• Overload protection: 600Vrms.
Содержание 72-2985
Страница 1: ...1 Mini Clamp Meter Model 72 2985 ...