3.7 Recall Data Operation

Under normal operating mode, push the 

RECALL

 button for retrieving the saved data from memory; follow the setting steps 

as below: 
1. Switch on the power supply
2. Push 

RECALL

 button to enter data recall mode, the minimum value position flashes and displays 

RECALL

 icon in LCD. 

3.   Press 

UP

 or 

DOWN

 arrow keys to recall the stored group number, the default values from voltage or current setting mode 

displayed in LCD. 

4. Press 

ENTER

 button to confirm data recall, or press 

STORE 

button to exit data recall mode.

    

                                

UP

      

 RECALL 

 1 

 2 

 ENTER

 --------> OK.

3.8 USB Power Charging 

Push the 

USB

 button, the default setting is voltage @ 5.2V and current @ 2.5A, the LCD will display 5.2V, with current value 

as 0A. Connect the mobile phone via USB cable for power charging or to supply USB power to any compatible device.
For battery charging, the USB ports are suitable for mobile phines. With automatic check function, the supply will set the 

proper charging current automatically. 
1. Push the 

OUT

 button to activate output mode. 

2. Push the 

USB 

button and 

USB

 indicator will turn off to exit USB charging mode.

 USB   OUT   

OK

4 Calibration 

Following the image below, connect a 5 digit displayed volt meter and current meter, please resistance (10Ω/100W) into the 

output terminals. To calibrate, start from point of zero voltage - voltage coefficient –and zero current – current coefficient. 

Hold the 

SET 

button to switch on power supply until “REF” is displayed in LCD to enter calibration mode.

4.1 Voltage Calibration

The power supply displays 

2.000A

 & 

05.00V

 in setting area of LCD and 

**.**V

 displayed in main part of LCD. Connect the 

output terminal with an external reference voltage meter and shift to CV mode. Hold the 

LEFT

 or 

RIGHT

 arrow keys to move 

the cursors to left or right and push the 

UP

 or 

DOWN

 arrow keys to adjust the values same as readings of external reference 

voltage meter, then click 

ENTER

 button to finish the voltage bias calibration. 

The power supply displays 

2.000A 

&

 30.00V

 in setting area of LCD, hold the 

LEFT

 or 

RIGHT

 arrow keys to move the cursors 

to left or  right and push the 

UP

 or 

DOWN

 arrow keys to adjust the values same as readings of an external reference voltage 

meter, then press 

ENTER

 button to finish the voltage gain calibration.

Содержание 72-2660

Страница 1: ...nell com www newark com www cpc co uk TM TM Page 1 V1 0 03 01 19 User Manual Switch Mode DC Power Supply Introduction Thank you for selecting the Tenma Switching DC Power Supply please read this user guide before the operation ...

Страница 2: ...pliance 72 2660 Switch Mode DC Power Supply is compliance with the specification described in this manual The content or specification of this manual is subject to change without prior notice Product Features The 72 2660 is the single output switch mode DC power supply with a maximum 30V output voltage 3 75A output current and maximum output power of 50W 72 2660 integrates the AC DC and DC DC 2nd ...

Страница 3: ...utor Power Cord 1PC User Guide 1PC Connect the power cord to the unit and to a properly grounded outlet and switch on 72 2660 The unit starts the self system checking the LCD displays 0 5s date of manufacturing production lot model number version number in turn 1 3 Quick Start OUT Button Press the OUT button to begin outputting the selected voltage and current The unit will display the voltage or ...

Страница 4: ... voltage can be adjusted from 0 3V to max 30V 1 4 2 The output constant current mode check This is for checking the functions of power supply under constant current mode 1 Switch on the product the power is off and the indicators of CC CV are off 2 Adjust voltage value to 30V 3 Connect the resistance 3Ω 50W between output terminals 4 Push the OUT button the indicator of CC also displays on the LCD...

Страница 5: ...f output offset Voltage 0 05 5mV Current 0 05 5mV Ripple and Noise 20HZ 20MHz Voltage 10mVrms 100mVp p Current 10mVrms 100mVp p Temperature Coefficient 0 40 C of output offset Voltage 0 05 Current 0 1 Dimensions 185mm 88mm 38mm Weight Net 370g 2 2 Supplementary Specifications Build in EEPROM Recommended Calibration Time Annually AC Input Power 90 265V AC 43 to 65Hz Operating Temperature 0 to 40 C ...

Страница 6: ...3 Push and SET ENTER button the max value position flashes of voltage setting area 4 Push the V A button the max value position of current setting flashes and current setting is activated 5 Push the LEFT or RIGHT arrow keys to move the cursors 6 Push the UP or DOWN arrow keys to change the settings 7 Push SET ENTER button and light off this button to exit current setting mode UP ENTER OUT SET V A ...

Страница 7: ...ly 1 Push the OUT button to activate output mode 2 Push the USB button and USB indicator will turn off to exit USB charging mode USB OUT OK 4 Calibration Following the image below connect a 5 digit displayed volt meter and current meter please resistance 10Ω 100W into the output terminals To calibrate start from point of zero voltage voltage coefficient and zero current current coefficient Hold th...

Страница 8: ...lting from negligence or where the Group was aware of the possibility of such loss or damage arising is excluded This will not operate to limit or restrict the Group s liability for death or personal injury resulting from its negligence Tenma is the registered trademark of the Group Premier Farnell Limited 2016 Part Number 72 2660 4 2 Current Calibration Push the V A button and power supply displa...

Отзывы: