Measuring DC Current

• 

Insert the black test lead into COM terminal. If the current under test is ≤ 200mA, 

insert the red test lead into the mA terminal. If the current is higher, insert the red 

test lead into 10A terminal.

• 

Turn the range selector to 

 range, and connect the test probes in series with 

the circuit to be measured. The polarity of the point touched by the red test probe 

together with the measured value is displayed.

Measuring AC Current

• 

Insert the black test lead into COM terminal. If the current under test ≤ 200mA, 

insert the red test lead into mA terminal. If the current is higher, insert the red test 

lead into 10A terminal.

• 

Turn the range selector to a 

~

 range, and connect the test probes with the circuit 

being measured in series.

Notes:

• 

If the current to be measured is unknown, choose the maximum measuring range 

(10A) and reduce it step by step until a satisfactory reading is obtained.

• 

Disconnect from the circuit under test when changing the range.

• 

If the display only shows “OL”, it indicates the measured current is out of range. In 

this case, turn the range selector to a higher range.

• 

“ ” on the meter indicates that the maximum input current is 200mA. The built-in 

fuse of 500mA/600V Φ6 x 32mm can effectively protect the circuit from burn-out. 

The 10A range uses a fuse of 10A/600V Φ6 x32mm for protection.

Measuring Resistance

• 

Insert the test leads into the input terminals (red to Ω and black to COM).

• 

Turn the range selector to a Ω range, and connect the test probes to the resistor 

being measured.

Notes

• 

If the measured resistance is out of the range selected, “OL” will be displayed and 

you should select a higher range. It normally takes several seconds to display a 

stable reading in a high-resistance (>1MΩ) measurement.

• 

When there is no input, such as in open-circuit condition, the instrument will 

display “OL”.

• 

When making an in circuit measurement, disconnect all power supplies and 

discharge all the capacitors.

• 

In 200MΩ range, a 10-digit reading is produced on a short circuit. It should be 

subtracted from the measured value. For instance, when you measure 100MΩ, the 

meter will display 101.0, subtract 10 digits from 101.0 to get the precise reading.

• 

Measuring Capacitance

• 

Before connecting to the capacitor to be tested, note that the zeroing after range 

change takes some time and the existing drift reading cannot influence the testing 

precision.

• 

Insert the test leads into the centre pair of terminals marked  .

• 

Turn the range selector to Fcx range, and connect the test probes to the capacitor 

under test.

• 

The capacitance is displayed.

Содержание 72-2590

Страница 1: ...1 Model 72 2590 and 72 2595 Digital Multimeter ...

Страница 2: ...0 seconds For current values over 5A the wait period between each measurement must be longer than 15 minutes When a measurement has been completed disconnect the testing probes from the circuit under test Replace the batteries as soon as the low battery indicator appears on the display Remove dead batteries from the meter or if it is not going to be used for a long time Never mix old and new batte...

Страница 3: ...bient temperature 23º 5º Relative Humidity 75 DC VOLTAGE Range Resolution Accuracy 200mV 100μV 0 5 1 2V 1mV 20V 10mV 200V 100mV 600V 1V 0 8 2 Note Input Impedance 10MΩ for all ranges AC VOLTAGE Range Resolution Accuracy 2V 1mV 0 8 3 20V 10mV 200V 100mV 600V 1V 1 5 5 Note Input Impedance 10MΩ for all ranges Frequency range 45Hz 400Hz Overload protection 600Vrms or 850Vpk pk Display Mean value RMS v...

Страница 4: ...sured voltage drop 200mV for full range AC CURRENT Range Resolution Accuracy 72 2590 72 2595 200μA 0 1μA 1 0 3 N A 2mA 1μA 1 0 3 N A 20mA 10μA N A 1 0 3 200mA 100μA 1 8 3 10A 10mA 3 0 5 Notes Overload Protection uA mA input 500mA 600V Φ6x32mm A input 10A 600V Φ6x32mm Maximum input current 10A For current over 5A measurement time must not exceed 10 seconds and the wait period between each measureme...

Страница 5: ... DC or AC for all ranges Relative Humidity 65 for measurements at 200MΩ range CAPACITANCE Range Resolution Accuracy 72 2590 72 2595 2nF 1pF N A 4 0 3 20nF 10pF N A 4 0 3 200nF 100pF N A 4 0 3 2μF 1nF 4 0 3 200μF 100nF 50μF 5 0 4 50μF for reference only Note Testing signal is about 200Hz 40mVrms TEMPERATURE 72 2595 ONLY Range Resolution Accuracy TEMP ºC 40ºC 1000ºC 40ºC 0ºC 1ºC 3 9 0ºC 400ºC 1 5 40...

Страница 6: ...or to range Connect the test probes with the power or load being measured in parallel and the polarity of the point touched by the red test probe displays B Measuring AC Voltage Insert the test leads into the input terminals red to V and black to COM Turn the range selector to a range and connect the test probes to the power or load being measured in parallel Notes If the voltage to be measured is...

Страница 7: ... in fuse of 500mA 600V Φ6 x 32mm can effectively protect the circuit from burn out The 10A range uses a fuse of 10A 600V Φ6 x32mm for protection Measuring Resistance Insert the test leads into the input terminals red to Ω and black to COM Turn the range selector to a Ω range and connect the test probes to the resistor being measured Notes If the measured resistance is out of the range selected OL ...

Страница 8: ...o range Connect the test probes with the diode being tested The reading obtained is the approximate value of the forward voltage drop Connect the test probes with the circuit under test if the resistance between the two tested points is 10Ω the built in buzzer sounds Battery Test 72 2590 only Insert the test leads into the input terminals red to and black to COM Turn the range selector to a range ...

Страница 9: ...d the power is off To replace the battery Unscrew and remove the battery cover Replace the battery with a new 9V battery of the same type Refit the cover Clean the meter with a clean soft cloth Do not use any chemicals abrasives or solvents that could damage the meter CLEANING Battery Screw ...

Страница 10: ...s is required Do not dispose of these items with general household waste Separate for the treatment recovery and recycling of the materials used Waste batteries can be returned to any waste battery recycling point which are provided by most battery retailers Contact your local authority for details of the battery and WEEE recycling schemes available in your area Made in China PR2 9PP Man Rev 1 0 ...