Please read these instructions carefully before use and retain for future 

reference.

This instrument is designed and manufactured in compliance with: G84793,  

lEC61010-1, CAT III 600V, Pollution Degree 2 and Double Insulation standards.

• 

Check the test leads, probes and case insulation before using. If you find any 

breakage or abnormality, or you consider the device is broken, stop using the 

device immediately.

• 

When using the test probes, keep your fingers behind the finger protection ring.

• 

Do not use the meter with the back cover open.

• 

Select appropriate test range for measurements.

• 

Ensure all inputs are less than the range selected otherwise it may cause electrical 

shock or meter damage.

• 

Do not change the range selector position during voltage or current 

measurements.

• 

Do not apply a voltage over 600V between COM terminal and ground.

• 

Take caution when working voltages are above 60V DC or 30V AC rms.

• 

Do not connect the meter to voltage signals when the range selector is on current, 

resistance, diode or continuity range.

• 

When measuring current, each single measurement should be shorter than 10 

seconds. For current values over 5A, the wait period between each measurement 

must be longer than 15 minutes.

• 

When a measurement has been completed, disconnect the testing probes from 

the circuit under test.

• 

Replace the batteries as soon as the low battery indicator appears on the display.

• 

Remove dead batteries from the meter or if it is not going to be used for a long 

time.

• 

Never mix old and new batteries together, or different types of batteries.

• 

Never dispose of batteries in a fire, or attempt to recharge ordinary batteries.

• 

Before replacing the battery, turn off the meter and disconnect all the test probes.

• 

To prolong battery life turn off the meter after use.

• 

CAT III: Measurement category III is for measurements performed in the building 

installation. Examples are measurements on distribution boards, circuit-breakers, 

and wiring, such as cables, bus-bars, junction boxes, switches and socket-outlets 

in the fixed installation, and equipment for industrial application and some other 

equipment such as stationary motors with permanent connection to the fixed 

installation.

• 

Only use test leads and probes that are in compliance with IEC 61010-031, and 

rated CAT III 600V.

SYMBOL GUIDE

Low battery 

Grounding

Warning

Continuity Buzzer

AC

DC

Fuse

Double insulated

Diode

IMPORTANT SAFETY INFORMATION

Содержание 72-2590

Страница 1: ...1 Model 72 2590 and 72 2595 Digital Multimeter ...

Страница 2: ...0 seconds For current values over 5A the wait period between each measurement must be longer than 15 minutes When a measurement has been completed disconnect the testing probes from the circuit under test Replace the batteries as soon as the low battery indicator appears on the display Remove dead batteries from the meter or if it is not going to be used for a long time Never mix old and new batte...

Страница 3: ...bient temperature 23º 5º Relative Humidity 75 DC VOLTAGE Range Resolution Accuracy 200mV 100μV 0 5 1 2V 1mV 20V 10mV 200V 100mV 600V 1V 0 8 2 Note Input Impedance 10MΩ for all ranges AC VOLTAGE Range Resolution Accuracy 2V 1mV 0 8 3 20V 10mV 200V 100mV 600V 1V 1 5 5 Note Input Impedance 10MΩ for all ranges Frequency range 45Hz 400Hz Overload protection 600Vrms or 850Vpk pk Display Mean value RMS v...

Страница 4: ...sured voltage drop 200mV for full range AC CURRENT Range Resolution Accuracy 72 2590 72 2595 200μA 0 1μA 1 0 3 N A 2mA 1μA 1 0 3 N A 20mA 10μA N A 1 0 3 200mA 100μA 1 8 3 10A 10mA 3 0 5 Notes Overload Protection uA mA input 500mA 600V Φ6x32mm A input 10A 600V Φ6x32mm Maximum input current 10A For current over 5A measurement time must not exceed 10 seconds and the wait period between each measureme...

Страница 5: ... DC or AC for all ranges Relative Humidity 65 for measurements at 200MΩ range CAPACITANCE Range Resolution Accuracy 72 2590 72 2595 2nF 1pF N A 4 0 3 20nF 10pF N A 4 0 3 200nF 100pF N A 4 0 3 2μF 1nF 4 0 3 200μF 100nF 50μF 5 0 4 50μF for reference only Note Testing signal is about 200Hz 40mVrms TEMPERATURE 72 2595 ONLY Range Resolution Accuracy TEMP ºC 40ºC 1000ºC 40ºC 0ºC 1ºC 3 9 0ºC 400ºC 1 5 40...

Страница 6: ...or to range Connect the test probes with the power or load being measured in parallel and the polarity of the point touched by the red test probe displays B Measuring AC Voltage Insert the test leads into the input terminals red to V and black to COM Turn the range selector to a range and connect the test probes to the power or load being measured in parallel Notes If the voltage to be measured is...

Страница 7: ... in fuse of 500mA 600V Φ6 x 32mm can effectively protect the circuit from burn out The 10A range uses a fuse of 10A 600V Φ6 x32mm for protection Measuring Resistance Insert the test leads into the input terminals red to Ω and black to COM Turn the range selector to a Ω range and connect the test probes to the resistor being measured Notes If the measured resistance is out of the range selected OL ...

Страница 8: ...o range Connect the test probes with the diode being tested The reading obtained is the approximate value of the forward voltage drop Connect the test probes with the circuit under test if the resistance between the two tested points is 10Ω the built in buzzer sounds Battery Test 72 2590 only Insert the test leads into the input terminals red to and black to COM Turn the range selector to a range ...

Страница 9: ...d the power is off To replace the battery Unscrew and remove the battery cover Replace the battery with a new 9V battery of the same type Refit the cover Clean the meter with a clean soft cloth Do not use any chemicals abrasives or solvents that could damage the meter CLEANING Battery Screw ...

Страница 10: ...s is required Do not dispose of these items with general household waste Separate for the treatment recovery and recycling of the materials used Waste batteries can be returned to any waste battery recycling point which are provided by most battery retailers Contact your local authority for details of the battery and WEEE recycling schemes available in your area Made in China PR2 9PP Man Rev 1 0 ...