ELECTRICAL SPECIFICATIONS

Current Source

Range

Display resolution

Accuracy

0-19.99mA

0.01mA

±(0.25%FS+1d)

0-24mA

0.1mA

±(0.5%FS+1d)

* Output 0 to 24mA current for loads up to 500ohms

* Output >20mA current for loads up to 400ohms

* FS: full scale

Current Measurement
Range

Display resolution

Accuracy

0-19.99mA

0.01mA

±(0.25%FS+1d)

0-24mA

0.1mA

±(0.5%FS+1d)

* FS: full scale

Power and Current Measurement of Two Wire Loop
Range

Display resolution

Accuracy

0-19.99mA

0.01mA

±(0.25%FS+1d)

0-24mA

0.1mA

±(0.5%FS+1d)

* Provides power DC 12V ±2V to the loop and measures current.

* FS: full scale

DC mV source
Range

Display resolution

Accuracy

-199.9mV to

+199.9mV

0.1mV

±(0.25%FS+1d)

* Output measured load impedance should >1K ohm.

* FS: full scale.

Note:

 the above specifications are tested under the environment RF field strength less 

than 3V/M and frequency less than 30MHz.

MEASURING PROCEDURE
Current Source

• 

Insert the black test lead into the negative (COM) jack and the red test lead into the 

positive (mV-mA) jack.

• 

Set the function switch to current (mA source) position.

• 

Press  the  0-19.9mA  button  to  set  the  0.01mA  display  resolution  (max  display  is 

19.99mA) or press the 0-24mA button to set the 0.1mA display resolution.

• 

Adjust the calibration rotary control (CAL ADJUST) to generate the current output.

Current Measurement

• 

Insert the black test lead into the negative (COM) jack and the red test lead into the 

positive (mV-mA) jack.

• 

Set the function switch to current (mA measure) position.

• 

Press  the  0-19.9mA  button  to  set  the  0.01mA  display  resolution  (max  display  is 

19.99mA) or press the 0-24mA button to set the 0.1mA display resolution.