background image

TM

TM

Page <9>

V1.0

27/05/21

Newark.com/exclusive-brands
Farnell.com/exclusive-brands
Element14.com/exclusive-brands

4 Beep 0=Off, 1=On

5 Lock 0=Lock, 1=Unlock

6 Output 0=Off, 1=On

7 N/A N/A
10. 

*IDN?

Description : The power supply identification returns

Example 

*IDN?

Contents TENMA 72‐2535 V2.0 (Manufacturer, model name,).

11. 

RCL<NR1>

Description : Recalls a panel setting.
NR1 1 – 5: Memory number 1 to 5
Example 

RCL1

 Recalls the panel setting stored in memory number 1

12. 

SAV<NR1>

Description : Stores the panel setting.
NR1 1 – 5: Memory number 1 to 5
Example : 

SAV1

 Stores the panel setting in memory number 1

13. 

OCP

< Boolean >

Description : Stores the panel setting.

Boolean: 0 OFF,1 ON

Example: 

OCP1

 Turns on the OCP

14. 

OVP

< Boolean >

Description : Turns on the OVP.

Boolean: 0 OFF,1 ON

Example: 

OVP1

 Turns on the OVP

Specifications

Note: The specifications in the table below have all been tested in temperatures ranging from 25°C down to -5°C, and after 

being warmed up for 20 minutes.

Part Number

72-10480/

72-2535

72-2540

72-2545/

72-2550

72-2925/

72-2930

72-2935/

72-2940

Voltage Range

0-30V

0-30V

0-60V

0-30V

0-60V

Current Range

0-3A

0-5A

0-2A (72-2545)

0-3A (72-2550)

0-10A

0-5A

Load Regulation

Voltage Current

≤0.01%+2mV

≤0.1%+5mA

≤0.1%+10mA

(72-2535)

≤0.01%+2mV

≤0.1%+5mA

≤0.01%+2mV

≤0.1%+5mA

≤0.01%+3mV

≤0.1%+20mA

≤0.01%+2mV

≤0.1%+10mA

Line Regulation

Voltage Current

≤0.01%+3mV

≤0.1%+3mA

≤0.01%+3mV

≤0.1%+3mA

≤0.01%+3mV

≤0.1%+3mA

≤0.01%+3mV

≤0.1%+3mA

≤0.01%+3mV

≤0.1%+3mA

Setup Resolution

Voltage Current

10mV

1mA

10mV

1mA

10mV

1mA

10mV

1mA

10mV

1mA

Содержание 72-10480

Страница 1: ...wark com exclusive brands Farnell com exclusive brands Element14 com exclusive brands Digital Control and Programmable DC Power Supply User Manual Part Number 72 2535 72 2540 72 2545 72 2550 72 2925 72 2930 72 2935 72 2940 72 10480 ...

Страница 2: ...semble unless you are qualified as service personnel AC Input AC input voltage 110V 120V 220V 230V 50 60Hz Connect the protective grounding conductor of the AC power cord to an earth ground in order to avoid electrical shock Operation Environment Location Indoor no direct sunlight dust free almost non conductive pollution Relative Humidity 80 Altitude 2000m Temperature 0 to 40 C Storage Environmen...

Страница 3: ...35 4 digit 4 digit Yes 10mV 1mA 72 2540 4 digit 4 digit Yes 10mV 1mA 72 2545 4 digit 4 digit Yes 10mV 1mA 72 2550 4 digit 4 digit Yes 10mV 1mA 72 2925 4 digit 4 digit No 10mV 1mA 72 2930 4 digit 4 digit Yes 10mV 1mA 72 2935 4 digit 4 digit No 10mV 1mA 72 2940 4 digit 4 digit Yes 10mV 1mA Performance Low noise cooling fan controlled by heatsink temperature Compact size light weight Operation Consta...

Страница 4: ... and the OVP indicator flickers Press the OVP key again and the power supply will recover OCP is OCP indicator When overcurrent function is turned on the OCP indicator light turns on C C is the constant current indicator When the power supply is in the mode of constant current the light will be on C V is the constant voltage indicator When the power supply is in the mode of constant voltage the li...

Страница 5: ...nect the AC power cord and select the corresponding AC voltage according to the back label then connect the AC power cord to the socket on the back panel Press the power switch to turn the power on The display initialises showing the model of the machine and then the setting level which is recalled from the last use Press the power switch again to turn the power off ...

Страница 6: ...ront panel key operation The key LED will turn on To unlock press and hold the LOCK key for two seconds Output Setup Panel Operation Connect the load to the front port CH1 Press the Voltage Current key to switch between the voltage adjustment and current adjustment Adjust the voltage and current with the Voltage Current adjustment knob By default the voltage and current knob work in coarse mode In...

Страница 7: ...lowing list Baud rate 9600 Parity bit None Data bit 8 Stop bit 1 Data fow control None Functionality Check Run this query command via the terminal application such as MTTTY Multithreaded TTY This should return the idenitifcation information Manufacturer model name serial number TENMA 72 2535 SN xxxxxxxx Vx xx Entering the Remote Control Mode Connect the USB The power supply will automatically conn...

Страница 8: ...oltage setting Example VSET1 Returns the CH1 voltage setting 5 IOUT X Description Returns the actual output current Example IOUT1 Returns the CH1 output current 6 VOUT X Description Returns the actual output voltage Example VOUT1 Returns the CH1 output voltage 7 BEEP Boolean Description Turns on or off the beep Boolean boolean logic Example BEEP1 Turns on the beep 8 OUT Boolean Description Turns o...

Страница 9: ...ting Boolean 0 OFF 1 ON Example OCP1 Turns on the OCP 14 OVP Boolean Description Turns on the OVP Boolean 0 OFF 1 ON Example OVP1 Turns on the OVP Specifications Note The specifications in the table below have all been tested in temperatures ranging from 25 C down to 5 C and after being warmed up for 20 minutes Part Number 72 10480 72 2535 72 2540 72 2545 72 2550 72 2925 72 2930 72 2935 72 2940 Vo...

Страница 10: ...mber 72 10480 72 2535 72 2540 72 2545 72 2550 72 2925 72 2930 72 2935 72 2940 Setup Accuracy 25 C to 5 C Voltage Current 0 5 20mV 0 5 5mA 0 5 20mV 0 5 10mA 0 5 30mV 0 5 5mA 0 5 20mV 0 5 20mA 0 5 30mV 0 5 10mA Ripple 20 20m Voltage Current 1mVrms 3mArms 2mVrms 3mArms 1mVrms 3mVrms 2mVrms 5mVrms 1mVrms 3mVrms Temperature Coeffcient Voltage Current 150ppm 150ppm 150ppm 150ppm 150ppm 150ppm 150ppm 150...

Отзывы: