Temperature Start
After started manually, the temperature condition device needs to meet to start recording, there are
3 temperature start types available for users to choose.
1.
Over,
device would start recording when temperature over the temperature value preset.
2.
Under,
device would start recording when temperature under the temperature value preset.
3.
Between,
device would start recording when temperature falls between the temperature
range preset.
Auto Start
Allows device to start recording on the specific date/time configured by user, logger would start recording
automatically when time is up, no need to press
Start/Stop
button after configuration.
Alarm Settings
Alarm Mode
No alarm setting
High & Low alarm
Multiple Alarm ranges, 5 alarm ranges available.*
*Tempod M1 does not support multiple alarm ranges.
Alarm Type
Single Event
Single event is that alarm could only be triggered when temperature over/below alarm limit for a
continuous period of time is not less than alarm delay.
Cumulative Alarm
Cumulative alarm means alarm could be triggered once the cumulative total time temperature over
or under alarm limit is not less than alarm delay.
NOTE
Only Tempod MP, Tempod MP-X of firmware version number not less than X1.6 support
temperature start capability.
Click
Write Configuration
button in down right corner of program to write in configuration parameters.