MM810A
16
17
Specifications (Cont.)
Temperature Coefficient: Nominal 0.15 x (specified accuracy) per °C
below 18 °C or above 28 °C
Automatic Power Off: After 30 minutes of inactivity.
To disable this feature, press
SELECT
while turning the meter on.
Noise Rejection*:
Normal Mode Rejection Ratio > 60 dB at 50 Hz and 60 Hz when measuring DCV
Common Mode Rejection Ratio > 60 dB from 0 Hz to 60 Hz when measuring ACV
Common Mode Rejection Ratio > 120 dB at 0 Hz, 50 Hz and 60 Hz when measuring DCV
Operating Conditions:
0 °C to 45 °C (32 °F to 113 °F), 0% to 80% relative humidity (non-condensing)
Altitude: 2000 m (6500') maximum
Indoor use only
Pollution Degree: 2
Storage Conditions: –20 °C to 60 °C (–4 °F to 140 °F)
0% to 80% relative humidity (non-condensing)
Remove battery.
Battery: 9-Volt (NEDA 1604, JIS 006P or IEC 6F22)
Overload Protections:
μA and mA: 0.44 A/1000 V DC/AC rms, interrupting rating 10 kA, F fuse,13/32" x 1-1/2"
A: 11 A/1000 V DC/AC rms, interrupting rating 20 kA, F fuse, 13/32" x 1-1/2"
V: 1100 V DC/AC rms
mV, Ω, and Other Functions: 1000V DC/AC rms
Safety: Double insulation per IEC/UL/EN61010-1 Ed. 3.0, IEC/EN61010-2-030 Ed. 1.0, IEC/
EN61010-2-033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 and CAN/CSA-C22.2 No. 61010-1-12
Ed. 3.0 to Category IV 1000 VAC and VDC
All Terminals: Category IV 1000 VAC and VDC
CENELEC DIRECTIVES: These meters conform to CENELEC Low-voltage directive 2006/95/EC
and Electromagnetic compatibility directive 2004/108/EC
* Noise rejection is the ability to reject unwanted signals, or noise.
• Normal mode voltages are AC signals that can cause inaccurate DC measurements. NMRR
(Normal Mode Rejection Ratio) is a measure of the ability to filter out these signals.
• Common mode voltages are signals present at the COM and + input terminals, with respect
to ground, that can cause digit rattle or offset in voltage measurements. CMRR (Common
Mode Rejection Ratio) is a measure of the ability to filter out these signals.
Statement of Conformity
Tempo Communications Inc. is certified in accordance with ISO 9001 (2015) for our Quality Management
Systems.
The instrument enclosed has been checked and/or calibrated using equipment that is traceable to the
National Institute for Standards and Technology (NIST).
22
Accuracy
(cont’d)
AC Current (AC + DC Current on DM-830A and DML-430A Only)
Range
Accuracy at
50 to 60 Hz
Accuracy at
40 to 1 kHz
Burden Voltage (typical)
(all frequency ranges)
600.0 µA
± (0.6% + 0.3 µA)
± (0.8% + 0.4 µA)
0.08 mV/µA
6000 µA
± (0.6% + 3 µA)
± (0.8% + 4 µA)
60.00 mA
± (0.6% + 0.03 mA)
± (0.8% + 0.04 mA)
2.1 mV/mA
600.0 mA
± (1.0% + 0.3 mA)
± (1.0% + 0.4 mA)
6.000 A
± (0.8% + 0.006 A)
± (0.8% + 0.006 A)
0.02 V/A
10.00 A
± (0.8% + 0.06 A)
± (0.8% + 0.06 A)
10 A continuous, 20 A for 30 sec. maximum with 5 minute cool down.
DC Current
Range
Accuracy at 50 to 60 Hz
Burden Voltage (typical)
(all frequency ranges)
600.0 µA
± (0.2% + 0.4 µA)
0.08 mV/µA
6000 µA
± (0.2% + 4 µA)
60.00 mA
± (0.2% + 0.04 mA)
2.1 mV/mA
600.0 mA
± (0.2% + 0.4 mA)
6.000 A
± (0.2% + 0.004 A
0.02 V/A
10.00 A
± (0.2% + 0.04 A)
10 A continuous, 15 A for 30 sec. maximum (20 A for DML-430A) with 5 minute cool down.
Line Level Frequency
Function
Sensitivity (Sine RMS)
Range
60.00 mV
40 mV
15.00 Hz to 50.00 kHz
600.0 mV
60 mV
15.00 Hz to 50.00 kHz
9.999 V
2.5 V
15.00 Hz to 10.00 kHz
99.99 V
25 V
15.00 Hz to 10.00 kHz
999.9 V
100 V
15.00 Hz to 10.00 kHz
600.0 µA
45 µA
15.00 Hz to 3.000 kHz
6000 µA
600 µA
15.00 Hz to 3.000 kHz
60.00 mA
40 mA
15.00 Hz to 3.000 kHz
600.0 mA
60 mA
15.00 Hz to 3.000 kHz
6.000 A
4 A
15.00 Hz to 3.000 kHz
10.00 A
6 A
15.00 Hz to 3.000 kHz