TIENet® Model 310 Ex Ultrasonic Level Sensor
Section 2 Installation and Setup for Signature
2-8
Figure 2-7 Sensor mounting options
2.4 Installation in
Hazardous Locations
Read all labels carefully before installing the equipment!
The TIENet Model 310 Ex device is ATEX-approved for use in
potentially explosive atmospheres when specific conditions are
met, as described in this section in reference to “X” Marking.
The 310 Ex is Group II, Category 1G equipment for use in gas
hazard zones 0, 1, and 2 (European standards), or Class I
Division 1 (North American standards).
The braid-drain lead depicted in Figure 2-8
Sensor labeling
regarding hazloc installations
is normally bonded to earth
through the Signature connector case terminals or conduit; it is
also electrically connected to the anti-static conductive housing
of the 310 Ex sensor.
Installation must be performed only by trained, qualified per-
sonnel.
Wall Mount
Approved for classified
hazardous locations.
Floor Stand
Not approved for classified
hazardous locations.
Suspension (cable straightener recommended)
Not approved for classified hazardous locations.
Содержание TIENet 310 Ex
Страница 2: ......
Страница 4: ......
Страница 14: ...TIENet Model 310 Ex Ultrasonic Level Sensor Section 1 Introduction 1 6...
Страница 35: ...TIENet Model 310 Ex Ultrasonic Level Sensor A 3...
Страница 36: ...TIENet Model 310 Ex Ultrasonic Level Sensor A 4...
Страница 38: ......
Страница 40: ......
Страница 42: ......