THCD-401 Instruction Manual
Page 19 of 47
(f)
Filtering Screen
This screen allows the user to modify the band and buffer size of the adaptive filter.
The filter only applies to the display and comms output readings. The band is represented as a percentage of
FS and can be set between 0.00% (OFF) and 1.00% and to ON. The buffer size allows settings between 0 and 6
seconds of filtering.
For more information on the adaptive filter, see the Principles section (7.6)
Switch Functions:
Func
Go back to Main Menu screen (in normal operation)
Undo changes (in edit mode)
◄
Move edit cursor left (in edit mode)
►
Move edit cursor right (in edit mode)
‘Wheel’
Move edit field selection (in normal operation)
Change char under cursor (in edit mode)
Ch1 to Ch4
No function
Open, Auto, Close
No function
Enter
Go to edit mode for selected field (in normal operation)
Accept changes (in edit mode)