Teledyne API T640 PM Mass Monitor
25
INLET INSTALLATION
1. To complete the installation now will depend on whether the instrument is being
installed in an Outdoor Enclosure or Walk-in Shelter. If performing testing in a
laboratory, you can proceed with attaching the inlet.
2. For the T640, assemble the inlet and the ASC as depicted below (Figure 2-13.
T640 Sample Head (5-LPM) and ASC Assembly).
Figure 2-13. T640 Sample Head (5-LPM) and ASC Assembly