Teledyne API - T360/T360M, 360E/360EM Appendix A Menu Trees (05233C DCN6552)
A-25
Name
1
Message Text
Description
WREARBOARD
REAR BOARD NOT DET
Rear board was not detected during
power up.
WRELAYBOARD
RELAY BOARD WARN
Firmware is unable to communicate with
the relay board.
WFRONTPANEL
FRONT PANEL WARN
Firmware is unable to communicate with
the front panel.
WANALOGCAL
ANALOG CAL WARNING
The A/D or at least one D/A channel has
not been calibrated.
Test Measurements
RANGE RANGE=50.0
PPM
3
CO RANGE=50.0 PPM
3, 7
D/A range in single or auto-range modes.
RANGE1 RANGE1=50.0
PPM
3
CO RANGE1=50.0 PPM
3, 7
D/A #1 range in dual range mode.
RANGE2 RANGE2=50.0
PPM
3
CO RANGE2=50.0 PPM
3, 7
D/A #2 range in dual range mode.
CO2RANGE
CO2 RANGE=20 %
7
CO
2
range.
O2RANGE
O2 RANGE=100 %
10
O
2
range.
STABILITY STABIL=0.0
PPM
3
CO STB=0.0 PPM
3, 7, 10
CO2 STB=0.0 %
7
O2 STB=0.0 %
10
Concentration stability (standard deviation
based on setting of
STABIL_FREQ
and
STABIL_SAMPLES
).
RESPONSE
2
RSP=0.20(0.00)
SEC Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
COMEAS
CO MEAS=4125.0 MV
Detector measure reading.
COREF
CO REF=3750.0 MV
Detector reference reading.
MRRATIO
MR RATIO=1.100
Measure/reference ratio.
AUTOZERO
4, 5
AZERO
RATIO=1.234
Measure/reference ratio during auto-
reference.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
PURGEPRESS
9
PURGE=7.5 PSIG
Purge pressure
VACUUM
8
VAC=6.8
IN-HG-A Vacuum pressure.
SAMPFLOW
6
SAMP FL=751 CC/M
Sample flow rate.
SAMPTEMP
13
SAMPLE TEMP=26.8 C
Sample temperature.
BENCHTEMP BENCH
TEMP=48.1
C Bench
temperature.
WHEELTEMP
WHEEL TEMP=68.1 C
Wheel temperature.
O2CELLTEMP
10
O2 CELL TEMP=50.2 C
O
2
sensor cell temperature.
BOXTEMP
BOX TEMP=26.8 C
Internal box temperature.
BOXTEMP2
4
BOX TEMP2=29.6 C
Internal box temperature #2.
OVENTEMP
11
OVEN TEMP=30.1 C
Oven temperature
PHOTOTEMP
PHT DRIVE=2500.0 MV
Photometer temperature.
07272B DCN6552
Содержание T360
Страница 2: ......
Страница 4: ...Model T360 T360M Operation Manual Teledyne API ii This page intentionally left blank 07272B DCN6552 ...
Страница 8: ...Model T360 T360M Operation Manual Teledyne API vi This page intentionally left blank 07272B DCN6552 ...
Страница 10: ...Model T360 T360M Operation Manual Teledyne API viii This page intentionally left blank 07272B DCN6552 ...
Страница 18: ...Teledyne API Model T360 T360M Operation Manual xvi This page intentionally left blank 07272B DCN6552 ...
Страница 188: ...This page intentionally left blank 07272B DCN6552 ...
Страница 264: ...07272B DCN6552 ...
Страница 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Страница 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Страница 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Страница 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Страница 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Страница 333: ...07272B DCN6552 D 7 ...
Страница 338: ...07272B DCN6552 D 12 ...