Teledyne API Model T360/T360M Operation Manual
Operating Instructions
149
4.13.8. Flow Calibration
The flow calibration allows the user to adjust the values of the sample flow rates
as they are displayed on the front panel and reported through COM ports to match
the actual flow rate measured at the sample inlet. This does not change the
hardware measurement of the flow sensors, only the software calculated values.
To carry out this adjustment, connect an external, sufficiently accurate flow meter
to the sample inlet (see Section 6.3.4 for more details). Once the flow meter is
attached and is measuring actual gas flow, press:
Adjust these values
until the displayed
flow rate equals the
flow rate being
measured by the
independent flow
meter.
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG ACAL DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
Exit at
any time
to return
to main
the
SETUP
menu
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Exit returns
to the
previous menu
Repeat Pressing
NEXT
until . .
.
DIAG
FLOW CALIBRATION
PREV NEXT
ENTR EXIT
DIAG FCAL
ACTUAL FLOW: 654 CC / M
0 6 5 4 ENTR EXIT
E
NTR
accepts the
new value and
returns to the
previous menu
EXIT
ignores the
new value and
returns to the
previous menu
Adjust these values until
the displayed flow rate
equals the flow rate being
measured by the
independent flow meter.
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
07272B DCN6552
Содержание T360
Страница 2: ......
Страница 4: ...Model T360 T360M Operation Manual Teledyne API ii This page intentionally left blank 07272B DCN6552 ...
Страница 8: ...Model T360 T360M Operation Manual Teledyne API vi This page intentionally left blank 07272B DCN6552 ...
Страница 10: ...Model T360 T360M Operation Manual Teledyne API viii This page intentionally left blank 07272B DCN6552 ...
Страница 18: ...Teledyne API Model T360 T360M Operation Manual xvi This page intentionally left blank 07272B DCN6552 ...
Страница 188: ...This page intentionally left blank 07272B DCN6552 ...
Страница 264: ...07272B DCN6552 ...
Страница 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Страница 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Страница 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Страница 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Страница 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Страница 333: ...07272B DCN6552 D 7 ...
Страница 338: ...07272B DCN6552 D 12 ...