Setup Mode Menus
105
5.9.2. ANALOG OUTPUT (DIAG AOUT)
Use Analog Output to verify functionality and accuracy of the analog outputs. The test
forces all analog output channels to produce signals ranging from 0% to 100% of the full
scale range in 20% increments. This test is useful to verify the operation of the data
logging/recording devices attached to the analyzer.
Section 12.7.6.1 presents instructions for use in troubleshooting and service.
5.9.3. ANALOG I/O CONFIGURATION (DIAG AIO)
The T200 analyzer comes equipped with four analog outputs. The first three outputs
(
A1 A2,
&
A3)
carry analog signals that represent the currently measured concentrating
of NO
x
, NO and NO
2
(see Section 5.4.2.1). The fourth output (
A4
) outputs a signal that
can be set to represent the current value of one of several test functions (see Table 5-9).
The following table lists the analog I/O functions that are available in the T200 analyzer.
Table 5-5: DIAG - Analog I/O Functions
SUB MENU
FUNCTION
MANUAL
SECTION
AOUT CALIBRATED
Initiates a calibration of the
A1
,
A2
,
A3
and
A4
analog output channels that
determines the slope and offset inherent in the circuitry of each output.
These values are stored and applied to the output signals by the CPU
automatically.
CONC_OUT_1
Sets the basic electronic configuration of the
A1
output (NO
x
Concentration).
There are four options:
RANGE
1
: Selects the signal type (voltage or current loop) and level of the
output
REC
OFS
: Allows them input of a DC offset to let the user manually adjust
the output level
AUTO CAL
: Enables / Disables the
AOUT CALIBRATION
Feature
CALIBRATED
: Performs the same calibration as
AOUT CALIBRATED
, but
on this one channel only.
CONC_OUT_2
Same as for CONC_OUT_1 but for analog channel
A2
(NO Concentration)
CONC_OUT_3
Same as for CONC_OUT_1 but for analog channel
A3
(NO
2
Concentration)
TEST OUTPUT
Same as for CONC_OUT_1 but for analog channel
A4
(TEST CHANNEL)
AIN CALIBRATED
Initiates a calibration of the A-to-D Converter circuit located on the Motherboard.
XIN1
.
.
.
XIN8
For each of 8 external analog inputs channels, shows the gain, offset,
engineering units, and whether the channel is to show up as a Test
function.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...