Setup Mode Menus
83
The T200 solves this problem by allowing the user to select a reporting range for the
analog outputs that only includes that portion of the physical range that covers the
specific application. This increases the reliability and accuracy of the analyzer by
avoiding additional gain-amplification circuitry.
Note
Only the reporting range of the analog outputs is scaled.
Both the DAS values stored in the CPU’s memory and the concentration values
reported on the front panel are unaffected by the settings chosen for the
reporting range(s) of the instrument.
5.4.2.1. ANALOG OUTPUT RANGES FOR NO
X
, NO AND NO
2
CONCENTRATION
The analyzer has three active analog output signals related to NO
x
, NO and NO
2
concentration, accessible through a connector on the rear panel.
NO
2
Concentration
ANALOG OUT
A1 A2 A3 A4
+ - + - + - + -
Test Channel
or O
2
concentration
(if optional O
2
sensor
is installed)
NO Concentration
NO
x
Concentration
Figure 5-1:
Analog Output Connector Pin Out
The
A1
,
A2
and
A3
channels output a signal proportional to the NO
x
, NO and NO
2
concentrations of the sample gas, respectively. The T200 can be set so that these outputs
operate in one of three modes: single range mode, independent range mode, or automatic
range mode (Section 5.4.3).
Additionally, the signal levels of
A1, A2
and
A3
outputs can be:
Configured full scale outputs of: 0 - 0.1 VDC; 0 – 1 VDC; 0 – 5 VDC or; 0 – 10 VDC
Equipped with optional 0-20 mADC current loop drivers (see Section 3.3.1.4 ) and
configured for any current output within that range analog output (e.g. 0-20 mA, 2-
20 mA, 4-20 mA, etc.)
Together these two sets of parameters allow the user a great deal of flexibility in how the
instrument reports NOx , NO and NO
2
concentration to external devices. For example,
using the
IND
mode the following configuration could be created:
A1
OUTPUT: NO
x
Output Signal = 4 – 20 mA representing 0-1000 ppb concentration
values
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...