
Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
160
5.3.2.6.
REMOTE ACCESS BY MODEM
The T100 can be connected to a modem for remote access. This requires a cable
between the analyzer’s COM port and the modem, typically a DB-9F to DB-25M cable
(available from Teledyne ML with P/N WR0000024).
Once the cable has been connected, check to ensure that the DTE-DCE is in the correct
position (refer to Section 3.4.8.3). Also ensure that the T100 COM port is set for a baud
rate that is compatible with the modem, which needs to operate with an 8-bit word
length with one stop bit.
The first step is to turn on the
MODEM ENABLE
communication mode (Mode 64,
Section 5.1.4). Once this is completed, the appropriate setup command line for your
modem can be entered into the analyzer. The default setting for this feature is:
AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0
This string can be altered to match your modem’s initialization and can be up to 100
characters long.
To change this setting, access the COMM Menu (refer to Figure 5-1), then press:
SETUP X.X
COM1 MODEM INIT:AT Y
&D
&H
<SET SET>
EDIT
EXIT
S
ETUP X.X COM1 MODEM INIT:
[A]
T Y
&D
&H
<CH CH> INS DEL [A]
ENTR
EXIT
SETUP X.X
COM1 MODE:0
SET>
EDIT
EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET
SET>
EDIT
EXIT
ENTR
accepts the
new string and returns
to the previous menu.
EXIT
ignores the new
string and returns to
the previous menu.
The
<CH
and
CH>
keys move
the
[ ]
cursor left and right
along the text string
The
INS
key
inserts a character
before the cursor
location.
Press the
[?]
button
repeatedly to cycle through
the available character set:
0-9
A-Z
space
’ ~ !
# $ % ^ & * ( ) - _ =
+[ ] { } < >\ | ; : , . / ?
The
DEL
key
deletes a character
at the cursor
location.
SETUP X.X
COMMUNICATIONS MENU
ID
COM1
COM2
EXIT
Select which
COM Port is
tested
Figure 5-26:
COMM
– Remote Access by Modem
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...