Command Descriptions
Each listing includes the command’s purpose, format, default setting (if applicable) range, Recom-
mended Setting, and description. When appropriate, we include amplifying notes and examples. If a
numeric value follows the command, the RiverPro/RioPro uses it to set a processing value (time, range,
percentage, processing flags). All measurement values are in metric units (mm, cm, and dm).
? – Help Menus
Purpose
Lists the major help groups.
Format
x? (see description)
Description
Entering
?
by itself displays all command groups. To display help for one command
group, enter
x?
, where x is the command group to view. When the RiverPro/RioPro dis-
plays the help for a command group, it also shows the format and present setting of those
commands. To see the help or setting for one command, enter the command followed by
a question mark. For example, to view the CB command setting, enter
CB?
.
Examples
See below.
>break
RiverPro
Teledyne RD Instruments (c) 2015
All rights reserved.
Firmware Version: 56.xx
?
Available Commands:
B ----------------------- Bottom Mode Commands
C ----------------------- Control Commands
E ----------------------- Environment Commands
M ----------------------- Loop Recorder Commands
O ----------------------- Feature Commands
P ----------------------- Performance Test Commands
S ----------------------- Sensor Control
T ----------------------- Time Commands
V ----------------------- Vertical Beam Mode Commands
W ----------------------- Water Profiling Commands
Y ----------------------- Display Banner
Z ----------------------- Vertical Beam Profile Commands
? ----------------------- Display Main Menu
>
>cb?
CB = 811 ----------------- Serial Port Control (Baud [8=115200]; Par; Stop)
>
The RiverPro/RioPro will hide menu options when in different water modes.
Vertical Beam commands are only available in 5 beam systems.