201/203/205/207 Series Flowmeters and Controllers Instruction Manual 

EAR99 Technology Subject to Restrictions Contained on the Cover Page

 

Page 22 of 28 

 

3.3.2. 

Miscellaneous Adjustments 

Periodically, during normal operation, the ZERO should be checked and adjusted when required. If the 
instrument is not shutting completely off when the Valve Override CLOSE function is active, or a 
command of zero flow has been given, the orifice may require approximately 1/8 turn clockwise. 

 

Endcap Removal 

The endcap on the inlet side must be removed to gain access to the filter or shunt assembly. First shut 
off the supply of gas to the instrument. Disconnect the Swagelok fittings on the inlet and outlet sides of 
the unit and remove it from the system plumbing. Remove the four hex bolts holding the endcap to the 
instrument. Carefully remove the endcap, filter, wave spring (if present) and shunt, noting their order and 
proper orientation. The shunt can be severely damaged if dropped. Examine the filter and shunt. If either is 
dirty or blocked, clean or replace as applicable. Reassembly is the reverse of the removal procedure. 
Recalibration of the HFC is necessary. 

 

Printed Circuit Board Replacement 

In the unlikely event that the PC board fails, it is easily removed from the instrument and replaced with a 
spare to minimize instrument downtime. Replacement of the PC board will require the instrument to be 
recalibrated. Unplug the power cable from the top of the transducer. Remove the two jackscrews next to the 

“D”

 connector and the two screws on the sides of the cover. Lift off the cover and unplug the four-wire sensor 

plug and the two-wire valve plug, noting their orientation prior to removal. Remove the screw that holds the 
PC board to the sensor. Troubleshoot or replace as applicable. Installation is the reverse of the above 
procedure. Recalibrate if any components were changed or if any potentiometers were adjusted or new. 

 

Sensor Replacement 

If the sensor fails or becomes plugged it can be removed. Remove the cover and the PC board per Section 3.5 
above. Remove the three bolts holding the sensor to the instrument base. Remove the sensor from the base 
noting the two O-rings between the sensor and the base. If the sensor is plugged it can be cleaned by running a 
fine wire (approximately 0.008" diameter) through the tube. If sensor needs replacement, obtain another from 
the factory and install it. Ensure that O-rings are clean and intact. Install O-rings on seating surface, then 
carefully place sensor over O-rings and tighten down the three screws evenly. Replacement of sensor will 
require recalibration per Section 3.3.1. 

 

Main Orifice Changes 

The orifice may require replacement if a large flow range change is desired, if changing to a gas that has a 
specific gravity significantly different than the original gas, if a large change in the differential pressures 
across the valve is desired or in the event that a small orifice becomes plugged. Replacement orifices can be 
acquired from the factory.  
 
The available main orifice diameters (inches) are 0.023, 0.032, 0.046, 0.086, 0.130, 0.156, 0.250, 0.312, 0.375, 
and 0.500. The half inch main orifices are only available for the HFC-203 models with the larger base fittings 

(3/4”) and for HFC

-207 instruments.  Hastings can offer orifices larger than half inch, but these items would 

need to be custom made. 
 
 
 

 

 

Содержание HASTINGS Everywhereyoulook 200 Series

Страница 1: ...CTION MANUAL www teledyne hi com I S O 9 0 0 1 C E R T I F I E D Use and Disclosure of Data Information contained herein is classified as EAR99 under the U S Export Administration Regulations Export reexport or diversion contrary to U S law is prohibited ...

Страница 2: ...tober 2009 Revision K Document Number 141 082010 August 2010 Revision L Document Number 141 072011 July 2011 Revision M Document Number 141 032021 March 2021 Visit www teledyne hi com for WEEE disposal guidance Description of Symbols and Messages used in this manual WARNING indicates a hazardous situation which if not avoided could result in death or serious injury Do not proceed beyond a WARNING ...

Страница 3: ... 2 Gas Quality 12 2 4 3 Upstream Differential Pressure 12 2 4 4 Other Mounting Fitting Options Notes 12 ELECTRICAL CONNECTIONS 13 2 5 1 Bipolar 15 VDC Power Supply to Bipolar Connections 13 2 5 2 Unipolar 24 VDC Power Supply to Unipolar Connections 14 2 5 3 General Connection Notes 14 OPERATION 15 2 6 1 Operating Conditions 15 2 6 2 Zero Check 15 2 6 3 High Pressure High Density Option 15 2 6 4 Ga...

Страница 4: ...ADJUSTMENTS 21 3 3 1 Calibration Procedure controller 21 3 3 2 Miscellaneous Adjustments 22 ENDCAP REMOVAL 22 PRINTED CIRCUIT BOARD REPLACEMENT 22 SENSOR REPLACEMENT 22 MAIN ORIFICE CHANGES 22 3 7 1 HFC 203 207 Orifice 23 4 WARRANTY AND REPAIR 24 WARRANTY REPAIR POLICY 24 RETURN POLICY 24 5 DRAWINGS 25 HFM 201 OUTLINE DRAWINGS 25 HFC 203 OUTLINE DRAWINGS 26 HFM 205 OUTLINE DRAWINGS 27 HFC 207 OUTL...

Страница 5: ... the use of moderate overpressures 500 psi 3 45 MPa or overflows Instruments are normally calibrated with the appropriate standard calibration gas nitrogen and a correction factor is used to adjust the output for the intended gas Special calibrations for other gases such as oxygen helium and argon are available upon special order NOTICE The instruments described in this manual may be available wit...

Страница 6: ... points MODULAR SENSOR WITH LOCKING CONNECTOR The HFM 201 203 205 207 Series incorporates a removable replaceable sensor module with a reliable sensor to board connection Field repairs to units can be achieved with a minimum of production line downtime with confidence in the sensor to PCB connection SETTLING TIME Filtering in the meter circuitry and more modern valve circuitry have improved the st...

Страница 7: ...C 20 mA 0 88 watts Power 15V meter 4 20 mA 14 16 VDC 55 mA 14 16 VDC 20 mA 1 2 watts Power 24V meter 0 5 VDC 11 36 VDC 1 08 watts Power 24V meter 4 20 mA 11 36 VDC 1 80 watts NOTICE Use of an underpowered or under voltage supply could result in equipment damage Output Flow Signal 0 5 00 VDC linear or 4 20 mA Wetted Material2 302 SS 316 SS Viton 82 18 Au Ni Braze Ni 200 Trace Silver Solder Connecto...

Страница 8: ...s Power 15V controller 4 20 mA 14 16 VDC 40 mA 14 16 VDC 175 mA 3 44 watts Power 24V controller 0 5 VDC 12 36 VDC 3 64 watts Power 24V controller 4 20 mA 13 36 VDC 3 96 watts NOTICE Use of an underpowered or under voltage supply could result in equipment damage Output Flow Signal CMD Input 0 5 00 VDC linear or 4 20 mA Wetted Material2 302 SS 316 SS Viton 82 18 Au Ni Braze Ni 200 Trace Silver Solde...

Страница 9: ...of DE 9 24 Volt The current loop is sourced internally from the 15 VDC supply and must be returned to the signal supply common after pin 5 15 VDC or pin 8 24 VDC passing through the monitoring circuitry to complete the loop The load must be between 0 and 600 Ohm The 4 20 mA I O option can accept a current input for flow controllers The 0 5 VDC command signal on pin 14 pin 3 for 24 Volts can be rep...

Страница 10: ...n off switch not be blocked Throughout this manual when reference is made to a power supply it is assumed the customer is using a Hastings power supply Designs and models offered as Hastings Power supplies change as new technology becomes available For knowledge of the most up to date Hastings power supplies please consult either Hastings sales staff or our website at http www teledyne hi com WARN...

Страница 11: ...er Requirements See specifications in sections 1 2 and 1 3 The HFM 200 HFC 202 Series supply voltages should be reasonably regulated as power supply ripple may propagate to the output Surge suppressors are recommended to prevent power spikes reaching the instrument The Hastings power supplies in Section 1 5 1 satisfy these requirements Hastings power supplies are limited to no more than 150VA max ...

Страница 12: ...125 0 31 mm so the instrument requires adequate filtering of the gas supply to prevent blockage or clogging of the tube 2 4 3 Upstream Differential Pressure The pressure regulator and the plumbing upstream must be of sufficient size to minimize changes in the upstream pressure In general high capacity regulators and large internal diameter plumbing help to make the system more stable When switchin...

Страница 13: ... 207 series flow instruments with anything other than the prescribed cables and power supplies discussed above can severely damage the instrument and void the warranty The figure below shows the schematic layout for connecting the instrument to an appropriate 15 Volt power supply A pinout is also given to show the signals available to the user and the signals given and used by the power supply Not...

Страница 14: ...of the DE 9 24 Volt is the command input This should be a 0 5 VDC or 4 20 mA signal and must be free of spikes or other electrical noise as these will generate false flow commands that the controller would attempt to adjust for Pin 15 of the DA 15 15 Volt and pin 1 of the DE 9 24 Volt is a regulated 5 VDC output 0 005 V reference which is occasionally used as a full scale command If a valve overri...

Страница 15: ...f not already energized Allow for a 1 hour warm up Stop all flow through the instrument and wait 2 minutes Caution Do not assume that all metering valves completely shut off the flow Even a slight leakage will cause an indication on the meter and an apparent zero shift For the standard 0 5 VDC output adjust the zero potentiometer located on the inlet side of the flowmeter until the meter indicates...

Страница 16: ...e flow rate of 100 SLM Controller B is a 5 V controller with a full scale flow rate of 10 SLM What gain factor on the output of controller A will result in controller B flowing 5 of controller A Use the formulas above 𝐺𝑎𝑖𝑛 𝐹𝑎𝑐𝑡𝑜𝑟 𝑜𝑓 𝐴 0 05 100 5 10 5 0 5 Cut the output of controller A in half and route the signal to the setpoint of controller B to flow 5 of controller A Operation with External Dev...

Страница 17: ...2 5 2 WARNING Care must be taken to avoid any high voltages that may be present when dealing with power supplies To successfully operate a flow controller a 0 5 VDC command signal proportional to the intended flow 0 Volts zero flow 5 Volts 100 of rated flow must be applied to pin 14 of the DA 15 15 Volt Pin 3 of the DE 9 24 Volt connector A 0 5 VDC signal proportional to the flow rate through the ...

Страница 18: ... on the left side of the pcb between the span potentiometer R7 and the large D connector J1 In the normal operating mode the jumper covers the left two pins To select an external sensor input move the jumper to the right two pins This swaps the flow input to the controller circuit from the flow meter output to pin 13 of the DA 15 connector 2 7 3 2 24 Volt Units For the 24 Volt units J8 is a six pi...

Страница 19: ...e to 100 and observe the flow through the controller If the overshoot is tolerable make no further changes Else remove the jumper and to prevent losing the unused jumper place it over only one pin of J15 one of the pins of J8 for 24 V boards Range Changing The range of the flow controller can be changed in the field if recalibration facilities are available The flow controller may require a differ...

Страница 20: ...s NOTICE adjusting the orifice too quickly or two far can damage the valve seat and cause leak paths that will not allow the controller to operate properly Please adjust the orifice gently and with caution SYMPTOM Output of unit is proportional to flow but extremely small and not correctable by span pot CAUSE Sensor is not being heated ACTION Unplug the sensor connector J4 for 15 V boards J5 for 2...

Страница 21: ...se of a calibration reference 1 Connect power to the instrument put in AUTO state and allow 10 flow for 30 minutes as a warm up 2 After the warm up period set the flow to zero and wait for 2 minutes Check to make sure there is zero flow 3 Adjust the ZERO potentiometer bottom hole on inlet side so that the flow instrument outputs 0 000 VDC 6 5 on 15V 2 8 on 24V 4 Override the valve closed or give a...

Страница 22: ... cover and unplug the four wire sensor plug and the two wire valve plug noting their orientation prior to removal Remove the screw that holds the PC board to the sensor Troubleshoot or replace as applicable Installation is the reverse of the above procedure Recalibrate if any components were changed or if any potentiometers were adjusted or new Sensor Replacement If the sensor fails or becomes plu...

Страница 23: ... The orifice is located at the bottom of the valve body and can be removed with a 9 16 socket wrench See the figure on right To re install an orifice first install the gasket onto the orifice replacement gaskets can be obtained from the factory Next screw the orifice into the valve base Snug up the orifice but do not overtighten Place diaphragm assembly into the base Line up the two small holes in...

Страница 24: ...rranty or non warranty repair must have been issued a Return Material Authorization RMA form RMA Forms may be obtained from the Information section of the Hastings Instruments website http www teledyne hi com inforeq htm WARNING Contaminated parts can be detrimental to health and environment Ensure instruments are free of hazardous contamination prior to shipment Company contact information TELEDY...

Страница 25: ...201 203 205 207 Series Flowmeters and Controllers Instruction Manual EAR99 Technology Subject to Restrictions Contained on the Cover Page Page 25 of 28 5 Drawings HFM 201 Outline Drawings ...

Страница 26: ...201 203 205 207 Series Flowmeters and Controllers Instruction Manual EAR99 Technology Subject to Restrictions Contained on the Cover Page Page 26 of 28 HFC 203 Outline Drawings ...

Страница 27: ...201 203 205 207 Series Flowmeters and Controllers Instruction Manual EAR99 Technology Subject to Restrictions Contained on the Cover Page Page 27 of 28 HFM 205 Outline Drawings ...

Страница 28: ...201 203 205 207 Series Flowmeters and Controllers Instruction Manual EAR99 Technology Subject to Restrictions Contained on the Cover Page Page 28 of 28 HFC 207 Outline Drawings ...

Отзывы: