37
64171
PS200
PORTABLE GAS MONITOR
5. Once gas is applied, press and hold the
R
button to start the calibration. The 60 second
countdown will commence (refer to
). After 60 seconds, the
monitor will be calibrated to the predefined values.
Figure 57: APPLY GAS Screen
6. The timer will count up to 9 seconds, during which time the monitor will adjust the gain to the
sensors accordingly. No manual input is required.
Figure 58: APPLY GAS Screen
7. Once calibration is successfully completed, the user has the option of saving the results (refer to
Figure 59: CAL COMPLETE/SAVE RESULTS Screen
).
Figure 59: CAL COMPLETE/SAVE RESULTS Screen
8.
Using the
L
button, highlight one of the 3 the available options :
A.
UPDATE – saves the calibration data in the memory and automatically updates the
calibration due date based on the option stored in the monitor set-up menu.
B.
SAVE ONLY – saves the calibration data in the memory but does not update the
calibration
due date.
C.
ABANDON – neither saves the calibration data in the memory nor updates the
calibration
due date.
9.
Press and hold the R button to accept this option
Revision 12
.1