PORTABLE GAS MONITOR
6.4. Quick/Full Bump Test
Operation now depends on whether QUICK or FULL bump is configured.
6.4.1 Quick Bump Test
As the alarm threshold for each range is exceeded, the audible/visual/vibration alarms will
activate and a
symbol will appear, otherwise a
symbol will be displayed.
6.4.2 Full Bump Test
After a short period of time, the gas readings are checked against configurable limits. The
audible/ visual/vibration alarms will activate and a
symbol will appear if the readings are
within these limits, otherwise a
symbol will be displayed.
6.5. Confirm Alarms
The user is prompted to confirm if the audible and visual alarms were activated, as shown in
Figure 45: Confirm Alarms
6.5.1 Bump Test – Pass
After selecting
YES
, BUMP TEST PASS is displayed, as shown in
. The
bump test result including date and time is automatically datalogged.
When the gas readings fall below their alarm set-points, or after 60 seconds, the bump test is
complete, and the monitor will automatically return to normal operation.
Figure 46: Bump Test Pass
Revision 12
.1