Oxygen Analyzer
Maintenance
Teledyne Analytical Instruments
51
4.4.2.1 S
PAN
F
AILURE
The analyzer checks the output of the cell at the end of the span. If
the raw output of the cell is less than 1.5 nA/ppb or more than 13.5
nA/ppb O
2
, the span will not be accepted. The analyzer will return to
the previous calibration values, and display at the top of the main
screen:
SPAN-CALIBRATION
ERROR
4.5
The
Alarms
Function
The Model BDS 3960 is equipped with 2 fully adjustable
concentration alarms and a system failure alarm. Each alarm has a relay
with a set of form “C" contacts rated for 3 amperes resistive load at 250
VAC. See Figure 3-5 in Chapter 3,
Installation
and/or the
Interconnection Diagram included at the back of this manual for relay
terminal connections.
The system failure alarm has a fixed configuration as described in
Chapter 3
Installation
.
The concentration alarms can be configured from the PPC software
as either
high
or
low
alarms by the operator. The alarm modes can be set
as
latching
or
non-latching
, and either
failsafe
or
non-failsafe
, or, they
can be
defeated
altogether. The setpoints for the alarms are also
established using this function.
Depending on your process, you can choose to configure the alarms
in a number of ways. Consider the following four points:
1. Which if any of the alarms are to be high alarms and which if
any are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the oxygen
concentration rises above the setpoint. Setting an alarm as LOW
triggers the alarm when the oxygen concentration falls below
the setpoint.
Decide whether you want the alarms to be set as:
•
Both high (high and high-high) alarms, or
•
One high and one low alarm, or
•
Both low (low and low-low) alarms.
Содержание BDS-3960
Страница 17: ...Oxygen Analyzer Introduction Teledyne Analytical Instruments 17 Figure 1 2 Model BDS 3960 Rear Panel...
Страница 29: ...Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 29...
Страница 64: ...Maintenance BDS 3960 Teledyne Analytical Instruments 64 Figure 5 3 Vacuum Degassing for the BDS Oxygen Sensor...
Страница 65: ...Oxygen Analyzer Maintenance Teledyne Analytical Instruments 65...