page 2

2. Precautions

Warranty limitation: The 8800P’s sample cell/piston assembly should not be disas-
sembled except by factory trained personnel. The warranty is voided, if the sample 
cell piston assembly is disassembled by the customer.

Avoid contaminated sample streams: The 8800P has been designed to allow for fast 
and precise measurements of dewpoints as low as -100 °C. Therefore, all surfaces in 
contact with the sample stream are made from electropolished stainless steel. This 
minimizes the time for the instrument to reach equilibrium with the sample stream, 
thus speeding up the measurement. The introduction of corrosive gases or gases 
contaminated with oils, solvents etc. may damage or coat the electropolished sur-
faces and slow down the measurement. Surfaces contaminated with oils etc. can be 
cleaned by factory trained personnel (please contact your sales representative). Sur-
faces damaged by corrosive gases are damaged permanently, causing longer 
response times. Gases containing chlorine, ammonia, HCl or SO

may also cause 

damage to the sensor. SO

can be monitored when the moisture content is low.

Avoid liquid water: Avoid the introduction of liquid water into the sample cell, as this 
may damage the sensor. Do not breathe directly into the instruments sample ports, 
as condensation may form which could damage the sensor element. If liquid water 
can not be avoided, order the XTR-65W sensor which is water proof, but has a limited 
range, and a slower dry down response.

Avoid high pressure: The 8800P has been designed for operation at pressures 
slightly above atmospheric, however each sample cell/sensor piston/dry-storage 
assembly has been tested for pressures up to 120 PSI. Do not introduce sample 
streams, which could cause pressure rises above 100 PSI. Make your measurements 
at atmospheric pressures, and use the pressure calculation feature to calculate dew-
points at higher pressures. Ask your representative for a regulator installation kit.

Exercise Caution with Hazardous Gases: Please use appropriate precautions if the 
sample you are about to measure is hazardous. Place the unit in a safe area, if it is to 
be used with hazardous gases. Please be aware that the possibility of a small leak 
always exists. Such a leak will not influence the accuracy of the measurement, but 
may represent a risk if the sample gas is of hazardous nature.

Avoid Extreme temperatures: Do not install the unit near heat sources such as radia-
tors or air ducts. The instruments’ operating temperature range is -10°C to 50°C.

Avoid Mechanical Vibration: Do not install the unit in places subject to extreme 
mechanical vibration or shock. Use resilient mounting, if shock and vibration are not 
avoidable. Call your representative if in doubt.

Avoid long exposures of sensor element to room air: For reasons explained later in 
this manual (section 7), do not expose the sensor to room air for longer then neces-
sary (1 - 2 Minutes). The sensor is exposed to room air, when the sensor actuator is 
in the out position and no sample line is connected, or the sample gas is shut off. It is 
best to push the sensor back into dry-storage, after a measurement is completed and 
whenever the instrument is not in use.

Do not force the Actuator: Remove any port obstructions (such as shipping caps) 
before operating the Actuator.

Содержание 8800P

Страница 1: ...instruction manual Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies Model 8800P TRACE MOISTURE ANALYZER...

Страница 2: ...presentative is Except as may be provided by contract this document and all specifica tions and drawings contained are the property of Teledyne Analytical Instruments are issued in strict confidence a...

Страница 3: ...siccant cartridge and lithium battery 2 2 VCO to Swagelok fittings and 2 Swagelok to barbed hose fittings 3 Calibration bulb 4 Shoulder strap Optional Equipment 5 VCO to VCO Fittings 6 Pig tail 7 Flex...

Страница 4: ...ressures slightly above atmospheric however each sample cell sensor piston dry storage assembly has been tested for pressures up to 120 PSI Do not introduce sample streams which could cause pressure r...

Страница 5: ...mple Cell 10 Electronic Board 5 Sensor 11 Instrument Case 6 Spring loaded PTFE seals The drawing on the left shows the 8800P with the sensor in dry storage position The sen sor is enclosed in a gas ti...

Страница 6: ...contamination with desiccant particles The factory supplied desiccant provides dry storage dewpoints as low as 80 C when fresh When the sensor is pushed into dry storage it will take some time to dry...

Страница 7: ...tubes are perfectly acceptable and offer advantages due to their flexibility for dew points above 65 C However PTFE tubes should be kept as short as possible Under no circumstances use rubber or plas...

Страница 8: ...ed if the dewpoint is decreasing Make sure the reading has completely stabilized before taking the final reading After the reading is taken push the sensor actuator back into the dry down position Not...

Страница 9: ...f the display Be careful not to change any parameter inadvertently 3 C F appear simultaneously to indicate the sensors attenuation in decibels 4 Values higher than 1999 and lower than 0 01 are display...

Страница 10: ...in and out of pressure cor rect mode When there is no pressure correction applied the PSI legend does not appear The PSI legend flashes at the bottom of the display when there is pressure cor rection...

Страница 11: ...splays the word 1 for 1 2 second and goes back to the Viewing Dewpoint Mode Autocal can be initiated by pressing the up button The unit will check if the sensor is in the sampling chamber if not it wi...

Страница 12: ...r dewpoint can be alternated in two different units 3 Selecting the sensor type In this mode the user can select the software matching the type of sensor installed in the instrument XTR 100 100 C to 2...

Страница 13: ...ared see the pin out table below In order to receive the dewpoint in the currently selected engineering units the host com puter must query the 8800P with C R The 8800P will respond with the dewpoint...

Страница 14: ...ull range of the sensor this feature may be useful in cases where a higher resolution output is required over a narrow dewpoint range or vise versa To verify or change the current loop scaling follow...

Страница 15: ...VDC to 25VDC when operating from external DC power The unit requires 12VACrms to 25VACrms when operating from external AC power In either case the unit can draw up to 50mA when the backlight is on and...

Страница 16: ...icro climate in sample cell while the display is flashing B 10 After one minute the instrument displays calibration numbers to be stored in the instruments memory displays 1 and thereafter goes back i...

Страница 17: ...e sensor actuator The replacement cartridge is in a sealed tin container Open the container and remove the replacement desiccant cartridge Remove the GREEN cap DO NOT open the RED cap as the desiccant...

Страница 18: ...to 50 C 231 Sensor circuit is open dewpoint displays tests check sensor wetup sensor to see if it will react 6 7 Sensor is saturated dry down the sensor 6 5 Sensor circuit is shorted check sensor Tryi...

Страница 19: ...PRESS LONG PRESS SHORT PRESS 1 2 sec LONG PRESS AND PSI MODE 1 minute few sec 1 sec OUT OK TOO LOW IN DESICCANT 65 20 CAL XSN 123 END SEN 2LO CNF AC AC _AC PUL OUT CAN SEN VALUE GAS VALUE F C DEWPOINT...

Страница 20: ...ve attenuation Manual calibration Set dewpoint corresponding to low end of analog output Factory default is low range of sensor e g 100 C for XTR 100 Factory default is high range of sensor e g 20 C f...

Страница 21: ...4000 6000 8000 10000 12000 14000 16000 18000 20000 22000 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 0 100 200 300 400 500 600 700 800 900 1000 1100 XTR100 4 20 mA output Instrument reading in ppmV...

Страница 22: ...3 0 04 0 05 0 06 0 07 0 08 0 09 0 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 1 2 3 4 5 6 7 8 9 10 20 30 Measured Dewpoint C Dewpoint 3 C Dewpoint 3 C Uncertainty of ppmV calculation due to 3 C measurment accur...

Страница 23: ...page 21 Appendix E Battery Life...

Отзывы: