page 14

 

7. Automatic Calibration

The instrument is calibrated at the factory with the shipped sensor, and does not need to 
be re-calibrated prior to use. 

Do not re-calibrate the unit unless you suspect a problem.

The 8800P instrument takes advantage of the pre-designed saturation point of Xentaur 
HTF sensors. Because Xentaur HTF sensors saturate at a factory designed dewpoint 
above +20°C, they can be calibrated by simply exposing them to a micro-climate with any 
dewpoint above +20°C and adjusting the calibration equation.
Remember that the micro-climate does not have to be accurate, it just has to be higher 
than the pre-designed saturation level. 
The automatic calibration procedure is as follows:

1. Make sure that the ambient temperature where you are performing the autocal is 

above +20°C. Check the temperature of the unit using its built in temperature indi-
cation.

2. Remove the adapter from the bulb (hand pump).
3. Moisten the sponge with 1 or 2 drops of water and replace the adapter back on 

the mouth of the bulb.

4. Turn instrument on. Refer to section 5.2.
5. Press the Mode key twice. The display will indicate 

&$/

6. Press the Up key, to confirm you want to perform a calibration.
7. Pull sensor actuator to measurement position, and pump into one of the input 

ports.

8. Push Up key to initiate the calibration procedure.
9. Keep pumping intermittently to assure wet micro climate in sample cell, while the 

display is flashing 

$&B$&

.

10. After one minute, the instrument displays calibration numbers to be stored in the 

instruments memory, displays 

(1'

 and thereafter goes back into measuring 

mode. The instrument is now calibrated.

11. Push in the sensor actuator.

Instrument calibration is recommended in 12 month intervals, the XTR65 sensor 

should be recalibrated after prolonged exposure to liquid water. Make sure the sensor 
has been sufficiently dried down and, simply follow steps 1 - 11 of the above procedure.

It is recommended to keep the sensor exposure to room air as short as possible to 

avoid super saturation of the sensor. While super saturation is not damaging to the sen-
sor, it will prolong the initial dry-down time after it is exposed to a dry environment. There-
fore, move the sensor out of the dry-storage only after you are ready to proceed with the 
calibration procedure and push the sensor in dry-storage immediately after the calibra-
tion procedure is completed. Under certain conditions, an over (super) saturated sensor 
may need to be completely dried out before either auto calibration or manual calibration 
is performed. Symptoms of these conditions are a sensor that will not go through the 
automatic calibration function to the 

(1'

display, or a sensor that will not dry down after 

calibration. To dry, install sensor in either a known dry gas stream i.e. instrument quality 
air or dry nitrogen, or push the sensor back into the dry-storage. After a minimum dry out 
period of 24 hours, proceed with the calibration procedure.

Содержание 8800P

Страница 1: ...instruction manual Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies Model 8800P TRACE MOISTURE ANALYZER...

Страница 2: ...presentative is Except as may be provided by contract this document and all specifica tions and drawings contained are the property of Teledyne Analytical Instruments are issued in strict confidence a...

Страница 3: ...siccant cartridge and lithium battery 2 2 VCO to Swagelok fittings and 2 Swagelok to barbed hose fittings 3 Calibration bulb 4 Shoulder strap Optional Equipment 5 VCO to VCO Fittings 6 Pig tail 7 Flex...

Страница 4: ...ressures slightly above atmospheric however each sample cell sensor piston dry storage assembly has been tested for pressures up to 120 PSI Do not introduce sample streams which could cause pressure r...

Страница 5: ...mple Cell 10 Electronic Board 5 Sensor 11 Instrument Case 6 Spring loaded PTFE seals The drawing on the left shows the 8800P with the sensor in dry storage position The sen sor is enclosed in a gas ti...

Страница 6: ...contamination with desiccant particles The factory supplied desiccant provides dry storage dewpoints as low as 80 C when fresh When the sensor is pushed into dry storage it will take some time to dry...

Страница 7: ...tubes are perfectly acceptable and offer advantages due to their flexibility for dew points above 65 C However PTFE tubes should be kept as short as possible Under no circumstances use rubber or plas...

Страница 8: ...ed if the dewpoint is decreasing Make sure the reading has completely stabilized before taking the final reading After the reading is taken push the sensor actuator back into the dry down position Not...

Страница 9: ...f the display Be careful not to change any parameter inadvertently 3 C F appear simultaneously to indicate the sensors attenuation in decibels 4 Values higher than 1999 and lower than 0 01 are display...

Страница 10: ...in and out of pressure cor rect mode When there is no pressure correction applied the PSI legend does not appear The PSI legend flashes at the bottom of the display when there is pressure cor rection...

Страница 11: ...splays the word 1 for 1 2 second and goes back to the Viewing Dewpoint Mode Autocal can be initiated by pressing the up button The unit will check if the sensor is in the sampling chamber if not it wi...

Страница 12: ...r dewpoint can be alternated in two different units 3 Selecting the sensor type In this mode the user can select the software matching the type of sensor installed in the instrument XTR 100 100 C to 2...

Страница 13: ...ared see the pin out table below In order to receive the dewpoint in the currently selected engineering units the host com puter must query the 8800P with C R The 8800P will respond with the dewpoint...

Страница 14: ...ull range of the sensor this feature may be useful in cases where a higher resolution output is required over a narrow dewpoint range or vise versa To verify or change the current loop scaling follow...

Страница 15: ...VDC to 25VDC when operating from external DC power The unit requires 12VACrms to 25VACrms when operating from external AC power In either case the unit can draw up to 50mA when the backlight is on and...

Страница 16: ...icro climate in sample cell while the display is flashing B 10 After one minute the instrument displays calibration numbers to be stored in the instruments memory displays 1 and thereafter goes back i...

Страница 17: ...e sensor actuator The replacement cartridge is in a sealed tin container Open the container and remove the replacement desiccant cartridge Remove the GREEN cap DO NOT open the RED cap as the desiccant...

Страница 18: ...to 50 C 231 Sensor circuit is open dewpoint displays tests check sensor wetup sensor to see if it will react 6 7 Sensor is saturated dry down the sensor 6 5 Sensor circuit is shorted check sensor Tryi...

Страница 19: ...PRESS LONG PRESS SHORT PRESS 1 2 sec LONG PRESS AND PSI MODE 1 minute few sec 1 sec OUT OK TOO LOW IN DESICCANT 65 20 CAL XSN 123 END SEN 2LO CNF AC AC _AC PUL OUT CAN SEN VALUE GAS VALUE F C DEWPOINT...

Страница 20: ...ve attenuation Manual calibration Set dewpoint corresponding to low end of analog output Factory default is low range of sensor e g 100 C for XTR 100 Factory default is high range of sensor e g 20 C f...

Страница 21: ...4000 6000 8000 10000 12000 14000 16000 18000 20000 22000 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 0 100 200 300 400 500 600 700 800 900 1000 1100 XTR100 4 20 mA output Instrument reading in ppmV...

Страница 22: ...3 0 04 0 05 0 06 0 07 0 08 0 09 0 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 1 2 3 4 5 6 7 8 9 10 20 30 Measured Dewpoint C Dewpoint 3 C Dewpoint 3 C Uncertainty of ppmV calculation due to 3 C measurment accur...

Страница 23: ...page 21 Appendix E Battery Life...

Отзывы: