Oxygen Analyzer
Part II: Analysis Unit
Part II: 3-5
Teledyne Analytical Instruments
Figure 3-3: Electrical Connector/Control Panel
For safe connections, ensure that uninsulated tips of the wires do not
extend beyond the terminal block screws to which they are attached.
Voltage Selector Switch:
Set the Voltage Selector switch to the source
voltage (115 or 230 V ac) that will be used to power the Analysis Unit
internal heater. Make sure the switch is set to the correct voltage BEFORE
making or energizing the power connections.
Power Connections:
115/230 V ac, 50/ 60 Hz power is required for
the heater that keeps the enclosure at a constant temperature. Connect per
standard power wiring codes. The connections are—
N
Neutral,
G
Ground,
H
Hot.
Fuse Installation:
Fuses are not installed at the factory. Be sure to
install the proper fuse (5
×
20 mm, 1.6 A) as part of installation. (See
Fuse
Replacement
in chapter 5,
Maintenance
.)
Solenoid and Sensor Signal Connections:
The Remote Probe con-
nector on the Control Unit (Part I, paragraph 3.3) connects to the Analysis
Unit's Solenoid Valves and Sensor Signal terminals. See Figure 3-4. It
provides signals to control the solenoid valves which regulate the zero, span
and sample gas flow, and accepts the sensor and thermistor signals for
processing.
Содержание 3010TB
Страница 8: ...viii Model 3010TB Teledyne Analytical Instruments...
Страница 12: ...iv Part I Model 3010TB Oxygen Analyzer Teledyne Analytical Instruments...
Страница 16: ...2 4 Part I 2 Operational Theory Model 3010TB Teledyne Analytical Instruments...
Страница 58: ...iv Part II Model 3010TB Oxygen Analyzer Teledyne Analytical Instruments...
Страница 62: ...1 4 Part II 1 Introduction Model 3010T Teledyne Analytical Instruments...
Страница 98: ...A 10 Appendix Model 3010TB Teledyne Analytical Instruments...