Paramagnetic O2 Purity Analyzer
Analysis Unit
Teledyne Analytical Instruments
73
Example:
Cross sensitivity calculation for a gas mixture:
The gas to be measured consists of 10 % CO
2
, 40 % N
2
, and 50% Ar.
Calculation of the zero value:
Table 6-2: Cross Sensitivity of Various Gases
Gas Cross
Gas Cross
Sensitivity
Sensitivity
in
vol.
%
in
vol.
%
Acetylene C
2
H
2
-0.24
Hydrogen chloride HCl
-0.30
Allene C
3
H
4
-0.44
Hydrogen Fluoride HF
+0.10
Ammonia NH
3
-0.26
Hydrogen Iodide HI
-1.10
Argon Ar
-0.22
Hydrogen Sulphide H
2
S
-0.39
Bromine Br
2
-1.30
Iodine
l
2
-2.40
1.2 Butadiene C
4
H
6
-0.49
Isobutane
C
4
H
10
-1.11
1.3 Butadiene C
4
H
6
-0.49
Isopentane
C
5
H
12
-1.49
n-Butane C
4
H
10
-1.11
Krypton
Kr
-0.51
l-Butane C
4
H
10
-0.85
Laughing gas N
2
O -0.20
cis 2- Butene C
4
H
8
-0.89
Methane
CH
4
-0.20
trans 2 Butene C
4
H
8
-0.92
Neon
Ne
+0.13
Carbon dioxide CO
2
-0.27
Neopentane
C
5
H
12
-1.49
Carbon monoxide CO
+0.06
Nitric acid HNO
3
+0.43
Chlorine Cl
2
-0.77
Nitrogen
dioxide NO
2
+28.00
137
0
:
ce
Interferen
Cross
Total
110
.
0
%
50
22
.
0
:
027
.
0
%
10
27
.
0
:
CO2
.
Ar
−
=
−
=
×
−
=
×
(
)
137
.
0
0.137
-
-
0%
Value
Point
Zero
Value
e
Inerferenc
Cross
-
0%O2
+
=
=
Содержание 3010MA
Страница 12: ...Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments xii...
Страница 14: ...Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 14...
Страница 24: ...Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 24...
Страница 64: ...Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 64...
Страница 66: ...Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 66...
Страница 82: ...Appendix Model 3010MA Teledyne Analytical Instruments 82 Approvals CE mark Area Classification General Purpose...