Part 1 Control Unit
Model 3010MA
Teledyne Analytical Instruments
32
36
System Alarm, normally closed contact
20
System Alarm, moving contact
37
System Alarm, normally open contact
Digital Remote Cal Inputs: Accept 0 V (off) or 24
V dc (on) inputs
for remote control of calibration. (See Remote Calibration Protocol
below.) See Table 3-3 for pin connections.
Zero: Floating input. 5 to 24 V input across the + and – pins
puts the analyzer into the
Zero
mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows
Zero
mode to terminate when done. A
synchronous signal must open and close the external zero valve
appropriately.
Span:
Floating input. 5 to 24 V input across the + and – pins
puts the analyzer into the
Span
mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows
Span
mode to terminate when done. A
synchronous signal must open and close external span valve
appropriately. See Figure 3-5 Remote Probe Connector.
(The –C option internal valves operate automatically.)
Cal
Contact: This relay contact is closed while analyzer is
spanning and/or zeroing. (See Remote Calibration Protocol
below.)
Table 3-3: Remote Calibration Connections
Pin Function
9
+ Remote Zero
11
–
Remote
Zero
10
+ Remote Span
12
–
Remote
Span
40
Cal
Contact
41
Cal
Contact
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 3010MA Analyzer, the customer's controller
must monitor the Cal Relay Contact.
Содержание 3010MA
Страница 12: ...Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments xii...
Страница 14: ...Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 14...
Страница 24: ...Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 24...
Страница 64: ...Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 64...
Страница 66: ...Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 66...
Страница 82: ...Appendix Model 3010MA Teledyne Analytical Instruments 82 Approvals CE mark Area Classification General Purpose...