24
58-723-004 Rev A00
Multiple
Output Triggers
The Kinetix camera has four independent trigger output signals. This enables hardware
control over light sources that cycle through different excitation wavelengths during a
sequence acquisition. For example, a multi-wavelength LED light source can rapidly switch
between two different fluorochromes.
The enabled outputs are simply cycled with each frame in a sequence. For example, if two
outputs are enabled, the output trigger signal is routed to output 1, output 2, output 1,
output 2, and so on.
If all four outputs are enabled the sequence would be as follows:
SMART
Streaming
Sequenced Multiple Acquisition Real Time Streaming, aka SMART Streaming, is an exclusive
Photometrics camera feature that enables Kinetix to capture a continuous sequence of
images, while cycling through a maximum of 16 pre-programmed exposure time values.
This avoids the overhead of host communication time, resulting in very high frame rate
imaging while maintaining the correct exposure level for each fluorophore.
The maximum exposure time per frame is 10 seconds in keeping with the SMART Streaming
high frame rate benefits.
When Multiple Output Triggers is combined with SMART streaming, it is possible to control
the exposure time of each output independently. This is much faster than using software-
based methods for control timing of illumination devices.
Trigger 1
Trigger 2
Trigger 3
Trigger 4
T
expose
Trigger 1
Trigger 2
Trigger 3
Trigger 4