7.2.2
Table No. 2
Isobutylene
Oxygen
(O
2
) 2 years
Oxygen
(O
2
) 1 year
Sensor P/N
6313998
6313780
6313817
Standard range (1)
0 – 1500ppm
isobutylene
2 - 30% volume
2 – 30% volume
Measurement principle
PID
Electrochemical
Electrochemical
Display resolution(1)
1ppm
0.1% v/v
0.1% v/v
Accuracy (2)
150
0.3% v/v
0.3% v/v
Repeatability (3)
0.1% v/v
0.1% v/v
Zero/Span drift (4)
0.2 / 2
0.2 / 2
Response time (5)
< 20s
< 10s
< 10s
Temperature (6)
0°C to +40°C
-20°C to +40°C
-20°C to +40°C
Relative humidity and
pressure range (7)
•
0 – 90% RH
•
1 bar ± 20%
•
10 – 95% RH
•
1 bar ± 20%
•
■
10 – 95% RH
•
1 bar ± 20%
Lifetime (8)
> 12 months
28 months
16 months
Storage conditions and
maximum storage time
(9)
•
4°C - 20°C
•
10-60% RH
•
1 bar ± 10%
•
6 months
maximum
•
4°C – 20°C
•
10 – 60% RH
•
1 bar ± 10%
•
3 months
•
■
4°C – 20°C
•
■
10 – 60% RH
•
■
1 bar ± 10%
•
■
3 months
Warm-up Delay (10)
60s
Operational upon start up
Notes
•
10.6ev lamp
•
Presence of high levels of CO2 can lead
to an over estimation of O2
concentration.
Содержание Everywhereyoulook BM 25A
Страница 1: ...NPB25AEN Revision D 0 BM 25A AW AREA GAS MONITOR USER MANUAL...
Страница 6: ...BM 25A AW AREA GAS MONITOR USER MANUAL VI NPB25AEN Revision D 0...
Страница 10: ...BM 25A AW AREA GAS MONITOR USER MANUAL 2 NPB25AEN Revision D 0 1 2 Product Overview Figure 2 Product Overview...
Страница 62: ...BM 25A AW AREA GAS MONITOR USER MANUAL 54 NPB25AEN Revision D 0...
Страница 67: ...BM 25A AW AREA GAS MONITOR USER MANUAL NPB25AEN Revision D 0 59 CSA Control Drawing for BM 25A AW Connections...
Страница 70: ...BM 25A AW AREA GAS MONITOR USER MANUAL 62 NPB25AEN Revision D 0...
Страница 71: ...BM 25A AW AREA GAS MONITOR USER MANUAL NPB25AEN Revision D 0 63...