OE695G Optical-to-Electrical Converter 

 

Reference Receiver 

The OE695G contains optional, built-in software reference receiver filters for 
common Fiber Channel, Ethernet, and ITU telecom standards. These 
reference receiver filters provide a 4-pole Bessel Thompson low pass filter 
response for the combined oscilloscope and optical-to-electrical (O-E) 
system with the -3dBe (electrical) power at 0.75* bitrate. 
Available reference receiver filters include: 

8GFC (8.5 Gb/s)  
10GBASE-W (9.953 Gb/s) 
OC192 (STM64) (9.953 Gb/s) 
10GBASE-R (10.3125 Gb/s) 
10GFC (10.519 Gb/s) 
ITU-T G.975 FEC (10.664 Gb/s) 

ITU-T G.709 FEC (10.709 Gb/s) 
10 GbE FEC (11.096 Gb/s) 
10GFC FEC (11.317 Gb/s) 
User Defined (up to 11.5 Gb/s) 
None (maximum 12.667 Gb/s)  

The software reference receiver supplied standard with the OE695G is based 
on a “nominal” oscilloscope response.  The typical uncertainty in this 

standard reference receiver frequency response is ±1.6 dBe up to Fref 
=0.75*bit rate and ±4 dBe 2*bitrate setting.  
An OE695G-REFCAL OE695G reference receiver calibration can be purchased 

that will provide less uncertainty with the OE695G when used with a specific 
oscilloscope and a specific channel.  This REFCAL is usually purchased at 
the time of initial OE695G and oscilloscope purchase.  If this is the case, 
Teledyne LeCroy will calibrate the specific OE695G to a specific serial 
number oscilloscope channel (channel 4 unless otherwise specified) and the 
resultant reference receiver uncertainty will be 

±

0.85 dBe up to Fref =0.75*bit 

rate, and ±4 dBe 2*bit rate setting on the specified oscilloscope input channel 

on the 11, 17, 20, 30, 39, 50, 75, 90, or 100 mV/div gain ranges.  The 
calibration will be indicated on the OE695G dialog, shown above.  Multiple 
OE695G can be calibrated to unique oscilloscope channels to allow usage of 
multiple OE695G.  
NOTE: If the OE695G-REFCAL is purchased after the initial 
OE695G/oscilloscope purchase, the oscilloscope must be returned to 
Teledyne LeCroy for the OE695G-REFCAL to be performed.   
A custom reference receiver (User Defined) can be applied for any bit rate up 
to 11.5 Gb/s.  Bit rates up to 12.5 Gb/s bit rates may be specified, but beyond 
11.5 Gb/s, the reference receiver requirements cannot be maintained.   

Содержание OE695G

Страница 1: ...Operator s Manual OE695G Optical to Electrical Converter...

Страница 2: ......

Страница 3: ...OE695G Optical to Electrical Converter Operator s Manual April 2017...

Страница 4: ...ly prohibited Customers are permitted to duplicate and distribute Teledyne LeCroy documentation for internal educational purposes Teledyne LeCroy is a registered trademarks of Teledyne LeCroy Inc Othe...

Страница 5: ...to be defective due to workmanship or materials and the defect is not caused by misuse neglect accident abnormal conditions of operation or damage resulting from attempted repair or modifications by a...

Страница 6: ...Optical Fiber 4 Operation with a Teledyne LeCroy Oscilloscope 4 Warning Messages 7 Performance Verification 8 CW Gain 8 Optical Reference Receiver Response 9 Care and Maintenance 11 Cleaning the OE695...

Страница 7: ...fety Instructions Do not install or terminate fibers while a light source is active Care must be taken to ensure that the instrument has been turned OFF before inspecting the end face s of the instrum...

Страница 8: ...ctions are all controlled from the oscilloscope s user interface The OE695G is unique in that it operates with a real time oscilloscope not a sampling oscilloscope and uses a digital signal processing...

Страница 9: ...eet reference receiver specification past 11 5 Gb s 62 5 125 m multi mode or single mode fiber input Broad wavelength range 750 to 1650 nm 7 dBm 5 mW max peak optical power Low noise as low as 25 pW H...

Страница 10: ...ected Use a torque wrench to ensure proper torque 6 Connect the other end of the LEMO power cable to the OE695G 7 To disconnect power pull on the LEMO connector tip to release the lock and pull out th...

Страница 11: ...e Receiver selection and Optical Wavelength selection functions Figure 1 OE695G tab of the oscilloscope s Channel dialog Dark Calibration The OE695G incorporates a Dark Calibration function to measure...

Страница 12: ...e receiver calibration can be purchased that will provide less uncertainty with the OE695G when used with a specific oscilloscope and a specific channel This REFCAL is usually purchased at the time of...

Страница 13: ...limited way If you see this message stop using the probe and contact Teledyne LeCroy for support Nominal reference receiver response applied The oscilloscope channel connected to the OE695G does not...

Страница 14: ...of the channel Enable measurement statistics 3 Connect the optical attenuator if used input fiber to the laser source or connect a fiber patch cable to the output of the laser source Before mating ca...

Страница 15: ...ptical attenuator Procedure 1 Connect the OE695G to a compatible oscilloscope channel with a channel specific calibration and allow to fully warm up 2 Preset the optical attenuator for high attenuatio...

Страница 16: ...OE695G optical input cover closed perform a dark calibration page 5 11 Connect the attenuator output cable to OE695G optical input 12 Configure the channel optical receiver for the desired response e...

Страница 17: ...connector adapter from the front panel to expose the internal fiber interface Calibration Interval The recommended calibration interval is one year Adjustment should be performed only by qualified pe...

Страница 18: ...ts case surrounded by the original packing material or equivalent 4 Label the case with a tag containing The RMA Name and address of the owner Product model and serial number Description of failure 5...

Страница 19: ...deos are available to help you get the most out of your TeledyneLeCroy products Go to teledynelecroy com support techlib The Datasheet published on the product page contains the detailed product speci...

Страница 20: ...s A 2 3 ELECTROMAGNETIC IMMUNITY IEC EN 61000 4 2 2009 Electrostatic Discharge 4 kV contact 8 kV air 4 kV vertical horizontal coupling planes 4 IEC EN 61000 4 3 A2 2010 RF Radiated Electromagnetic Fie...

Страница 21: ...demonstrated to the following specifications as listed in the Official Journal of the European Communities IEC EN 61010 031 2015 Safety requirements for electrical equipment for measurement control an...

Страница 22: ...OE695G Optical to Electrical Converter 16...

Страница 23: ......

Страница 24: ...928329 00 RevA April 2017...

Отзывы: